Reducing False Alarms and Alarm Fatigue from Pulse Oximeters in a Neonatal Care Unit: A Quality Improvement Study

Nawaz Z. Patel, Dipen Vasudev Patel, Ajay Gajanan Phatak, Vinit G. Patel, Somashekhar Marutirao Nimbalkar

Journal of Neonatology · 2022-06 · 36 (2) · 135-142

DOI: 10.1177/09732179221100531 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Neonatology, online erschienen am 02.06.2022. Verfasst von Nawaz Z. Patel und 4 weiteren Autor:innen. Die Arbeit wurde bisher 14-mal zitiert (Crossref) und nennt 21 Literaturangaben.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Pulse oximeters play a vital role in monitoring sick newborns, but majority of the times, their alarms are of little clinical significance. False alarms from pulse oximeters were frequent in our neonatal intensive care unit, causing desensitization of staff and inattention to clinically important alarms.

Objectives

The objective of the study was to reduce the burden of false alarms and alarm fatigue.

Methods

This was a quality improvement (QI) study. During the baseline phase, alarms for heart rate and SPO 2 (percentage of oxygenated blood) were evaluated. Response to them within 30 s was checked. Major causes of false alarms were the movement of the newborns, inappropriate alarm limits, and loose probe. Fishbone chart was prepared, and Plan-Do-Study-Act (PDSA) model was designed using point of care quality improvement (POCQI) principles. During the first PDSA (implementation phase), the staff was trained to adjust alarm limits using the reference chart. A lecture on noise hazards, emphasis on muting the alarms, and advice to change probe site at beginning of their duty was given. In the second PDSA (sustenance phase), components of the first PDSA were strengthened by supportive supervision and guidance and the training of new staff.

Results

Total 494, 329, and 162 alarms were observed during baseline, implementation, and sustenance phases, respectively. In the baseline phase majority, 364 (73.7%) were false alarms, which occurred at a frequency of 0.39 alarms/patient/h. More than one-third of the alarms were not responded timely. After QI, false alarms reduced to 214 (65%; 0.22 alarms/patient/h) and 100 (61.7%; 0.16 alarms/patient/h) in implementation and sustenance phases, respectively. Response to true alarms improved from 101 (77.7%) to 104 (90.4%) to 62 (100%) during the 3 phases consecutively.

Conclusion

Frequent false alarms lead to desensitization of health care workers. POCQI principles can be a practical approach for reducing false alarms and improving alarm safety.

Zitieren

APA

Patel, N. Z., Patel, D. V., Phatak, A. G., Patel, V. G., & Nimbalkar, S. M. (2022). Reducing False Alarms and Alarm Fatigue from Pulse Oximeters in a Neonatal Care Unit: A Quality Improvement Study. Journal of Neonatology, 36(2), 135-142. https://doi.org/10.1177/09732179221100531

Vancouver

Patel NZ, Patel DV, Phatak AG, Patel VG, Nimbalkar SM. Reducing False Alarms and Alarm Fatigue from Pulse Oximeters in a Neonatal Care Unit: A Quality Improvement Study. Journal of Neonatology. 2022;36(2):135-142. doi:10.1177/09732179221100531

BibTeX
@article{Patel2022,
  author = {Patel, Nawaz Z. and Patel, Dipen Vasudev and Phatak, Ajay Gajanan and Patel, Vinit G. and Nimbalkar, Somashekhar Marutirao},
  title = {Reducing False Alarms and Alarm Fatigue from Pulse Oximeters in a Neonatal Care Unit: A Quality Improvement Study},
  journal = {Journal of Neonatology},
  year = {2022},
  volume = {36},
  number = {2},
  pages = {135--142},
  doi = {10.1177/09732179221100531}
}
Kostenlos starten →