Multi-scale topology optimisation design and mechanical property analysis of porous interbody fusion cage
Bio-Medical Materials and Engineering · 2025
DOI: 10.1177/09592989241303291 ↗Kurz gefasst
Fachartikel in der Zeitschrift Bio-Medical Materials and Engineering, 2025. Verfasst von Huaiyue Zhang und 2 weiteren Autor:innen. Beteiligte Einrichtung: School of Mechanical Engineering, Xinjiang University, Urumqi 830017, China. Hauptthemen laut MEDLINE: Spinal Fusion, Lumbar Vertebrae. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 44 Literaturangaben. Gefördert von: The National Natural Science Foundation of China.
Abstract (Originalsprache)
Background
Titanium (Ti) and polyether ether ketone (PEEK) interbody fusion cages cause postoperative stress shielding problems. The porous cage design is one of the solutions advanced to mitigate this problem.
Objective
Exploring the mitigation of stress shielding with a porous interbody fusion cage after surgery for idiopathic scoliosis.
Methods
The porous interbody fusion cage was constructed based on the multiscale topology optimisation method, and the postoperative lumbar spine models implanted with it. The porous Ti and PEEK fusion cages were evaluated under physiological conditions to investigate their mechanical properties.
Results
The volume of the porous fusion cage was reduced by 52.57%, and the stress was increased by 242.76% and 252.46% compared with the Ti and PEEK fusion cage; the modulus of elasticity of the porous fusion cage was reduced by 76.85%, and the strain was increased by 131.40%∼686.51% compared with the Ti cage; the porous fusion cage increased L3 cortical bone stress by 13.36% and 13.52% and cancellous bone by 82.93% and 76.72%, respectively, compared with the original interbody fusion cages.
Conclusion
The porous interbody fusion cage has a much more lightweight design which facilitates growth of bone tissue. However, a frame structure should be constructed to minimize issues with stress peaks and localised stress concentrations. It also has a significantly lower stiffness which helps alleviate vertebral stress shielding, further fostering bone growth. The porous fusion cage thus meets the clinical requirements for better fusion outcomes.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- School of Mechanical Engineering, Xinjiang University, Urumqi 830017, China.
- The Sixth Affiliated Hospital, Xinjiang Medical University, Urumqi 830002, China.
ORCID-Profile: Rongchang Fu
Förderung
- The National Natural Science Foundation of China
Zitieren
APA
Zhang, H., Fu, R., & Zhu, X. (2025). Multi-scale topology optimisation design and mechanical property analysis of porous interbody fusion cage. Bio-Medical Materials and Engineering. https://doi.org/10.1177/09592989241303291
Vancouver
Zhang H, Fu R, Zhu X. Multi-scale topology optimisation design and mechanical property analysis of porous interbody fusion cage. Bio-Medical Materials and Engineering. 2025. doi:10.1177/09592989241303291
BibTeX
@article{Zhang2025,
author = {Zhang, Huaiyue and Fu, Rongchang and Zhu, Xu},
title = {Multi-scale topology optimisation design and mechanical property analysis of porous interbody fusion cage},
journal = {Bio-Medical Materials and Engineering},
year = {2025},
doi = {10.1177/09592989241303291}
}
EUVIMED