The effect of tilt degree and position changes on video ocular counter-roll test performance
Journal of Vestibular Research · 2025
DOI: 10.1177/09574271251382787 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Vestibular Research, 2025. Verfasst von Asya Fatma Men und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Audiology, Hamidiye Faculty of Health Sciences, Hamidiye University of Health Sciences, Istanbul, Turkey. Hauptthemen laut MEDLINE: Eye Movements, Reflex, Vestibulo-Ocular, Posture. Die Arbeit nennt 29 Literaturangaben.
Abstract (Originalsprache)
Background
The vestibular system is essential for balance and visual stability and relies on the ocular counter-roll (OCR) reflex to regulate torsional eye movements during head tilt.
Objective
This study aimed to investigate how different tilt angles (15° and 45°) and tilt positions (head and body) influence static video Ocular Counter-Rolling (vOCR) responses in healthy individuals.
Methods
A total of 30 healthy volunteers (17 females and 13 males), aged between 18 and 30 years, participated in the study. vOCR testing was conducted using the Interacoustics VisualEyes™ 525 video-oculography (VOG) system. Torsional eye movements were recorded at 15° and 45° during both lateral head tilt and body block tilt. Torsion degree, asymmetry, and gain values were analyzed and compared across all conditions.
Results
Gain values did not significantly differ based on tilt angle or position (p > 0.05). The degree of torsional eye movement was significantly higher at 45° tilts compared to 15°, and greater during body tilt than head tilt (p < 0.0001). Asymmetry values remained within normal limits under all conditions and showed no significant differences (p > 0.05).
Conclusions
These findings indicate that the angle and position of tilt affect the magnitude of the rotational response, but that gain and asymmetry values remain consistent across conditions. This suggests that comparable gains can be obtained at different angles, such as 15° and 45°, and that low-angle tilts may be sufficient for evaluating otolith-ocular gain.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Audiology, Hamidiye Faculty of Health Sciences, Hamidiye University of Health Sciences, Istanbul, Turkey.
ORCID-Profile: Asya Fatma Men
Zitieren
APA
Men, A. F., & Küçük Ceyhan, A. (2025). The effect of tilt degree and position changes on video ocular counter-roll test performance. Journal of Vestibular Research. https://doi.org/10.1177/09574271251382787
Vancouver
Men AF, Küçük Ceyhan A. The effect of tilt degree and position changes on video ocular counter-roll test performance. Journal of Vestibular Research. 2025. doi:10.1177/09574271251382787
BibTeX
@article{Men2025,
author = {Men, Asya Fatma and Küçük Ceyhan, Ayşenur},
title = {The effect of tilt degree and position changes on video ocular counter-roll test performance},
journal = {Journal of Vestibular Research},
year = {2025},
doi = {10.1177/09574271251382787}
}
EUVIMED