Biomechanical evaluation of a fully cortical-threaded screw for modified cortical bone trajectory fixation: Combined experimental and finite element study
Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine · 2026-07 · 240 (7) · 663-675
DOI: 10.1177/09544119261461660 ↗Kurz gefasst
Fachartikel in der Zeitschrift Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, online erschienen am 28.06.2026. Verfasst von Yixi Wang und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Minimally Invasive Spine and Precision Orthopedics, The First Affiliated Hospital of Xinjiang Medical University, Urumqi, China. Hauptthemen laut MEDLINE: Bone Screws, Cortical Bone, Spinal Fusion. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 59 Literaturangaben. Gefördert von: Xinjiang Uygur Autonomous Region Tianshan Talents Medical and Health High-Level Personnel Training Program.
Abstract (Originalsprache)
Although fully cortical-threaded screws are introduced as a design adaptation tailored to the cortical-dominant load path of modified cortical bone trajectory (MCBT) fixation, the biomechanical consequences of this design across the screw–bone interface and fusion construct remain insufficiently defined. Therefore, this study compared fully cortical-threaded MCBT screws with clinically used dual-threaded screws through an integrated experimental and finite element (FE) framework, spanning in vitro ovine vertebral biomechanical test and specimen-specific L4 vertebra and L1–S1 fusion models. At the screw level, compared with the control group, fully cortical-threaded MCBT screws increased maximum pull-out strength by 111.2% and multidirectional stiffness by 75%–89% in L4 vertebral FE analysis, with biomechanical testing showing corresponding increases of 39.1% in insertion torque and 41.9% in maximum pull-out strength. At the fusion-construct level, fully cortical-threaded MCBT fixation limited fused-segment motion and decreased stress across the cage, instrumentation, and vertebral bone, indicating a more coordinated load-transfer pattern rather than a simple increase in interface-level strength. These effects were consistent across fusion strategies, but procedure-specific mechanics remained evident, with PLIF producing more symmetric load sharing and TLIF retaining intrinsic asymmetry owing to unilateral facet joint resection. Overall, the fully cortical-threaded screw design for MCBT promoted continuous and stable bone–screw load transfer, translating interface-level gains into coordinated load distribution, greater fusion-construct stability, and lower deformation-driven stress concentration. These findings indicate that aligning screw architecture with the cortical-dominant load path is a mechanically rational design strategy within MCBT fixation, particularly in biomechanically demanding settings.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Minimally Invasive Spine and Precision Orthopedics, The First Affiliated Hospital of Xinjiang Medical University, Urumqi, China.
- Department of Orthopedics, Jingzhou First People's Hospital, China.
ORCID-Profile: Paerhati Rexiti
Förderung
- Xinjiang Uygur Autonomous Region Tianshan Talents Medical and Health High-Level Personnel Training Program
Zitieren
APA
Wang, Y., Xiao, Y., Yang, J., Zhang, R., Chen, Q., & Rexiti, P. (2026). Biomechanical evaluation of a fully cortical-threaded screw for modified cortical bone trajectory fixation: Combined experimental and finite element study. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 240(7), 663-675. https://doi.org/10.1177/09544119261461660
Vancouver
Wang Y, Xiao Y, Yang J, Zhang R, Chen Q, Rexiti P. Biomechanical evaluation of a fully cortical-threaded screw for modified cortical bone trajectory fixation: Combined experimental and finite element study. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine. 2026;240(7):663-675. doi:10.1177/09544119261461660
BibTeX
@article{Wang2026,
author = {Wang, Yixi and Xiao, Yang and Yang, Jitao and Zhang, Rui and Chen, Qihao and Rexiti, Paerhati},
title = {Biomechanical evaluation of a fully cortical-threaded screw for modified cortical bone trajectory fixation: Combined experimental and finite element study},
journal = {Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine},
year = {2026},
volume = {240},
number = {7},
pages = {663--675},
doi = {10.1177/09544119261461660}
}
EUVIMED