EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC 4.0

Radiolucent lines in a cementless hip stem: The use of semi-automated measurements to monitor progression over a 10 year period

Graham H Isaac iD, Rosyln S Cassidy, Janet C Hill, Gregory M Pryce, Sophie Williams, Ruth K Wilcox, David E Beverland

Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine · 2026-07 · 240 (7) · 641-649

DOI: 10.1177/09544119261451379 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, online erschienen am 09.06.2026. Verfasst von Graham H Isaac und 6 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Medical and Biological Engineering, School of Mechanical Engineering, University of Leeds, UK. Hauptthemen laut MEDLINE: Hip Prosthesis, Arthroplasty, Replacement, Hip. Die Arbeit nennt 12 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: This work was funded by DePuy Synthes, Warsaw, Indiana, USA..

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Radiolucent lines (RLLs) observed adjacent to a cementless stem following Total Hip Arthroplasty (THA) have been linked to stem loosening. Their manual assessment is susceptible to observer error. In this study a semi-automated (Auto) method was developed to characterise RLLs from Anterior-Posterior (AP) and Lateral (Lat) radiographs, evaluate their time progression and effects on clinical outcomes. The study comprised of 237 subjects with postoperative, 1 and 10 Y radiographs. Of these 131 also had 4–7 Y radiographs. All subjects also had accompanying implant, demographic and clinical outcomes data. An algorithm was written to evaluate the incidence of RLLs and post-processing was performed to characterise RLLs as ‘not present’ or ‘benign’, or significant (‘SigRLL’). The criteria for SigRLL were tuned to minimise the number of false negatives compared to a manual characterisation. The incidence of SigRLL using AP/Lat or AP only was compared and the latter was used to monitor their development over time. The relationship with clinical outcomes was evaluated. Incidence of SigRLLs using the Auto and Manual AP/Lat methods agreed in 199/237 (84.0%) of subjects. The discrepancies included seven false negatives. The Auto AP/Lat and AP methods agreed in 221/237 (93.2%) of subjects. Twenty-eight subjects with a 4–7 Y review had SigRLLs at 10 Y. Of these, 23 had SigRLLs at 4–7 Y. Greater pain was reported in the SigRLLs group ( p = 0.003). Four out of five stems subsequently revised for loosening exhibited 10 Y SigRLLs.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Hip ProsthesisArthroplasty, Replacement, HipHumansFemaleAutomationMaleMiddle AgedAgedProsthesis FailureRadiographyTime Factors

Stichwörter der Autor:innen

cementless hipsfemoral stem loosinghip revisionradiographsradiolucent linestotal hip arthroplasty

Autor:innen und Institutionen

  • Institute of Medical and Biological Engineering, School of Mechanical Engineering, University of Leeds, UK.
  • Primary Joint Unit, Musgrave Park Hospital, Belfast, UK.

ORCID-Profile: Graham H Isaac

Förderung

  • This work was funded by DePuy Synthes, Warsaw, Indiana, USA.

Zitieren

APA

Isaac, G. H., Cassidy, R. S., Hill, J. C., Pryce, G. M., Williams, S., Wilcox, R. K., & Beverland, D. E. (2026). Radiolucent lines in a cementless hip stem: The use of semi-automated measurements to monitor progression over a 10 year period. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 240(7), 641-649. https://doi.org/10.1177/09544119261451379

Vancouver

Isaac GH, Cassidy RS, Hill JC, Pryce GM, Williams S, Wilcox RK, et al. Radiolucent lines in a cementless hip stem: The use of semi-automated measurements to monitor progression over a 10 year period. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine. 2026;240(7):641-649. doi:10.1177/09544119261451379

BibTeX
@article{Isaac2026,
  author = {Isaac, Graham H and Cassidy, Rosyln S and Hill, Janet C and Pryce, Gregory M and Williams, Sophie and Wilcox, Ruth K and Beverland, David E},
  title = {Radiolucent lines in a cementless hip stem: The use of semi-automated measurements to monitor progression over a 10 year period},
  journal = {Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine},
  year = {2026},
  volume = {240},
  number = {7},
  pages = {641--649},
  doi = {10.1177/09544119261451379}
}
Kostenlos starten →