Human factors testing in a novel automated peritoneal dialysis system
Peritoneal Dialysis International: Journal of the International Society for Peritoneal Dialysis · 14.08.2026
DOI: 10.1177/08968608261476545 ↗Kurz gefasst
Fachartikel in der Zeitschrift Peritoneal Dialysis International: Journal of the International Society for Peritoneal Dialysis, online erschienen am 14.08.2026. Verfasst von Nupur Gupta und 5 weiteren Autor:innen. Beteiligte Einrichtung: Division of Nephrology, Indiana University School of Medicine, Indianapolis, IN, USA. Die Arbeit nennt 21 Literaturangaben. Gefördert von: Simergent.
Abstract (Originalsprache)
Background
Because automated peritoneal dialysis (APD) systems are used in the home and operated by patients or their care partners, the human–device interface must be thoughtfully designed to facilitate intuitive operation to reduce the risk of errors that could occur during their use. User-friendly design should also help reduce anxiety to peritoneal dialysis (PD) adoption by patients incident to end-stage kidney disease (ESKD). Usability issues remain a significant barrier to PD adoption and are an important contributor to premature death, serious injuries, and PD technique failure.
Methods
A summative human factors usability study was conducted on a novel, gravity-based APD device ( Archimedes ™) with 15 current or former PD patients and 15 dialysis nurses. Participants were trained for 2 h, followed by a training decay period, then evaluated with critical Use Scenario tasks consisting of Simulated Use tasks and Knowledge tasks reflecting tasks that could result in patient harm if performed incorrectly.
Results
Of the Simulated Use tasks evaluated, 97.3% were deemed successful across all users. This high success rate demonstrates the effectiveness of the Archimedes APD device in facilitating the tasks required for PD. For patients, success outcomes were achieved in 96.8% of Simulated Use tasks. Nurses achieved a success rate of 97.8% for Simulated Use tasks evaluated. For Knowledge tasks, success outcomes were achieved in 99.3% and 98.6% of tasks for patients and nurses.
Conclusion
In this summative human factors usability study, the Archimedes APD system usability was found to be safe, accessible and easy for its intended users and use environments after the relatively short training period compared to the status quo.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Division of Nephrology, Indiana University School of Medicine, Indianapolis, IN, USA.
- Division of Nephrology and Hypertension, Northwestern University Feinberg School of Medicine, Chicago, IL, USA.
- Division of Kidney Diseases and Hypertension, The George Washington University, Washington, DC, USA.
- Simergent LLC, Oklahoma City, OK, USA.
ORCID-Profile: Nupur Gupta, Vikram Aggarwal, James A Sloand
Förderung
- Simergent
Zitieren
APA
Gupta, N., Aggarwal, V., Sloand, J. A., Pendergraft, R., Lindo, S. J., & Miller, B. (2026). Human factors testing in a novel automated peritoneal dialysis system. Peritoneal Dialysis International: Journal of the International Society for Peritoneal Dialysis. https://doi.org/10.1177/08968608261476545
Vancouver
Gupta N, Aggarwal V, Sloand JA, Pendergraft R, Lindo SJ, Miller B. Human factors testing in a novel automated peritoneal dialysis system. Peritoneal Dialysis International: Journal of the International Society for Peritoneal Dialysis. 2026. doi:10.1177/08968608261476545
BibTeX
@article{Gupta2026,
author = {Gupta, Nupur and Aggarwal, Vikram and Sloand, James A and Pendergraft, Rick and Lindo, Steve J and Miller, Brent},
title = {Human factors testing in a novel automated peritoneal dialysis system},
journal = {Peritoneal Dialysis International: Journal of the International Society for Peritoneal Dialysis},
year = {2026},
doi = {10.1177/08968608261476545}
}
EUVIMED