An in vitro comparison of bacterial touch contamination prevention between a novel APD cycler connector and existing APD connectors

James A Sloand iD, Jeremy P Burton, Chris I Munday, Arsh K Jain, Osama El Shamy, Rick Pendergraft, Steve J Lindo

Peritoneal Dialysis International: Journal of the International Society for Peritoneal Dialysis · 23.06.2026

DOI: 10.1177/08968608261461078 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Peritoneal Dialysis International: Journal of the International Society for Peritoneal Dialysis, online erschienen am 23.06.2026. Verfasst von James A Sloand und 6 weiteren Autor:innen. Beteiligte Einrichtung: Kidney Diseases & Hypertension, The George Washington University School of Medicine & Health Sciences, Washington, USA. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 24 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Peritoneal dialysis (PD)—associated peritonitis remains a leading cause of hospitalization, technique failure, and transfer to in-center hemodialysis, with touch contamination during connections accounting for many episodes. Although shrouded connectors and “flush before fill” mechanisms reduce contamination risk, most automated peritoneal dialysis (APD) systems incorporate recessed fluid paths on only one side of the tubing–bag interface, potentially allowing bacterial transfer. We evaluated whether bilateral shrouded connectors with recessed fluid paths reduce bacterial contamination during simulated APD therapy.

Methods

In vitro microbiological testing was performed using three APD systems: Vantive/Baxter Claria™ (System A), Fresenius Liberty™ (System C), and Simergent Archimedes™ (System B). Connectors were deliberately contaminated with Staphylococcus aureus to simulate touch contamination, followed by a simulated APD treatment. Bacterial growth was quantified from pre-therapy flush fluid and the first patient fill (“mock peritoneum”). Agar imprinting was used to assess bacterial distribution relative to connector geometry.

Results

Bacterial growth adjacent to the fluid path was observed with contaminated System A connectors but not with System B or System C, where growth was confined to the shroud. Quantitative cultures demonstrated markedly higher bacterial transfer with System A compared with System B and System C across both flush and mock peritoneal samples. Systems B and C employing bilateral shrouded, recessed connectors showed bacterial counts several orders of magnitude lower. All aseptic controls were sterile.

Conclusion

Bilateral shrouded connectors with recessed fluid paths substantially reduce transfer of touch contaminants during simulated APD therapy. This design may mitigate a major mechanism of PD-related peritonitis, warranting clinical evaluation.

Schlagwörter

Stichwörter der Autor:innen

Automated peritoneal dialysisdialysis connector designinfection controlperitonitis preventionshrouded connectorstouch contamination

Autor:innen und Institutionen

  • Kidney Diseases & Hypertension, The George Washington University School of Medicine & Health Sciences, Washington, USA.
  • Simergent LLC, Oklahoma City, OK, USA.
  • St. Joseph's Health Care, Lawson Research Institute, London, Ontario, Canada.
  • Department of Microbiology and Immunology and the Department of Surgery (Division of Urology), Faculty at Western University, London, Ontario, Canada.
  • Department of Medicine (Division of Nephrology) and the Department of Epidemiology & Biostatistics, Faculty at Western University, London, Ontario, Canada.
  • Department of Medicine, University of Pennsylvania, Philadelphia, PA, USA.

ORCID-Profile: James A Sloand

Zitieren

APA

Sloand, J. A., Burton, J. P., Munday, C. I., Jain, A. K., Shamy, O. E., Pendergraft, R., & Lindo, S. J. (2026). An in vitro comparison of bacterial touch contamination prevention between a novel APD cycler connector and existing APD connectors. Peritoneal Dialysis International: Journal of the International Society for Peritoneal Dialysis. https://doi.org/10.1177/08968608261461078

Vancouver

Sloand JA, Burton JP, Munday CI, Jain AK, Shamy OE, Pendergraft R, et al. An in vitro comparison of bacterial touch contamination prevention between a novel APD cycler connector and existing APD connectors. Peritoneal Dialysis International: Journal of the International Society for Peritoneal Dialysis. 2026. doi:10.1177/08968608261461078

BibTeX
@article{Sloand2026,
  author = {Sloand, James A and Burton, Jeremy P and Munday, Chris I and Jain, Arsh K and Shamy, Osama El and Pendergraft, Rick and Lindo, Steve J},
  title = {An in vitro comparison of bacterial touch contamination prevention between a novel APD cycler connector and existing APD connectors},
  journal = {Peritoneal Dialysis International: Journal of the International Society for Peritoneal Dialysis},
  year = {2026},
  doi = {10.1177/08968608261461078}
}
Kostenlos starten →