Grating based laboratory 5.41 keV X-ray phase contrast microscopy with zone-plate: design, instrumentation and validation
Journal of X-Ray Science and Technology · 2026
DOI: 10.1177/08953996261464117 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift Journal of X-Ray Science and Technology, 2026. Verfasst von Yuhang Tan und 7 weiteren Autor:innen. Beteiligte Einrichtung: Research Center for Advanced Detection Materials and Medical Imaging Devices, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences…. Hauptthemen laut MEDLINE: Microscopy, Phase-Contrast. Die Arbeit nennt 26 Literaturangaben. Gefördert von: Shenzhen Science and Technology Program, National Natural Science Foundation of China, National Key Research and Development Program of China.
Abstract (Originalsprache)
Background
It has been demonstrated that grating interferometers can be utilized to enhance the X-ray imaging contrast of low-density and weakly absorbing objects.
Objective
In this study, the design, instrumentation and validation of a grating based laboratory 5.41 keV X-ray phase-contrast microscope with zone-plate are reported.
Methods
This dedicated microscope was carefully designed by coupling the zone-plate imaging method and the grating interferometer imaging method. Specially fabricated X-ray optics were integrated along with a certain X-ray tube source illumination module, motion stages and X-ray detection module. The system’s spatial resolution was characterized with the modulation transfer function (MTF) analysis and the power spectral density (PSD) analysis. The absorption-contrast and phase-contrast images were extracted from the acquired phase stepping dataset. The splitting and superposition behavior of the phase-contrast signal was evaluated by comparing the experimental results with the numerical simulation. CT projections of a mitochondria-targeted, antibody-stained Jurkat T cell were pre-processed by geometric registration before performing the standard filtered back-projection (FBP) reconstruction.
Results
The absorption imaging results of the Siemens star phantom demonstrated that this microscope achieved a spatial resolution of 35.0 nm based on MTF analysis, while the minimum detectable structure size was 28.5 nm based on PSD analysis. The phase-contrast imaging of a 5.00 μ m diameter polystyrene microsphere demonstrated the splitting phenomenon ( 8.54 μ m on the sample plane) of the ± 1 st order components, and both the splitting distance and line profiles agreed well with the theoretical imaging model. In the reconstructed phase-contrast CT images, the mitochondria could be clearly identified.
Conclusions
This study reported the development of a grating based laboratory 5.41 keV X-ray phase-contrast microscope with zone-plate. Results showed that it could significantly enhance the imaging performance of low-density objects such as polystyrene microsphere and biological cells.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Research Center for Advanced Detection Materials and Medical Imaging Devices, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, Shenzhen, Guangdong, China.
- Paul C Lauterbur Research Center for Biomedical Imaging, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, Shenzhen, Guangdong, China.
- State Key Laboratory of Biomedical Imaging Science and System, Shenzhen, Guangdong, China.
- Research Center for Medical Artificial Intelligence, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, Shenzhen, Guangdong, China.
ORCID-Profile: Yongshuai Ge
Förderung
- Shenzhen Science and Technology Program
- National Natural Science Foundation of China
- National Key Research and Development Program of China
Zitieren
APA
Tan, Y., Yang, J., Zhu, J., Zhang, X., Zheng, H., Liang, D., Li, Z., & Ge, Y. (2026). Grating based laboratory 5.41 keV X-ray phase contrast microscopy with zone-plate: design, instrumentation and validation. Journal of X-Ray Science and Technology. https://doi.org/10.1177/08953996261464117
Vancouver
Tan Y, Yang J, Zhu J, Zhang X, Zheng H, Liang D, et al. Grating based laboratory 5.41 keV X-ray phase contrast microscopy with zone-plate: design, instrumentation and validation. Journal of X-Ray Science and Technology. 2026. doi:10.1177/08953996261464117
BibTeX
@article{Tan2026,
author = {Tan, Yuhang and Yang, Jiecheng and Zhu, Jiongtao and Zhang, Xin and Zheng, Hairong and Liang, Dong and Li, Zhicheng and Ge, Yongshuai},
title = {Grating based laboratory 5.41 keV X-ray phase contrast microscopy with zone-plate: design, instrumentation and validation},
journal = {Journal of X-Ray Science and Technology},
year = {2026},
doi = {10.1177/08953996261464117}
}
EUVIMED