Inclusion of spatio-energetic charge sharing effect model for accurate photon counting CT simulation
Journal of X-Ray Science and Technology · 2025
DOI: 10.1177/08953996251323725 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of X-Ray Science and Technology, 2025. Verfasst von Jiabing Sheng und 7 weiteren Autor:innen. Beteiligte Einrichtung: Guangdong Provincial Key Laboratory of Medical Image Processing, School of Biomedical Engineering, Southern Medical University, Guangzhou, China. Hauptthemen laut MEDLINE: Tomography, X-Ray Computed. Die Arbeit nennt 37 Literaturangaben. Gefördert von: Key Technologies Research and Development Program, Guangzhou Municipal Science and Technology Project, Science and Technology Planning Project of Shenzen Municipality.
Abstract (Originalsprache)
Background
Photon counting CT has demonstrated exceptional performance in spatial resolution, density resolution, and image quality, earning recognition as a groundbreaking technology in medical imaging. However, its technical implementation continues to face substantial challenges, including charge sharing effects.
Objective
To develop a spatio-energetic charge-sharing modulation model for accurate photon counting CT simulation (SmuSim). Specifically, SmuSim is built upon the previously developed photon counting toolkit (PcTK) and thoroughly incorporates the charge sharing effects that occur in photon counting CT.
Methods
The proposed SmuSim firstly enrolls three primary modules, i.e., photon transport, charge transport, and charge induction to characterize the charge sharing effects in the photon counting CT imaging chain. Then, Monte Carlo simulation is also conducted to validate the feasibility of the proposed SmuSim with well-built charge sharing effects model.
Results
Under diverse detector configurations, SmuSim's energy spectrum response curves exhibit a remarkable alignment with Monte Carlo simulations, in stark contrast to the Pctk results. In both digital and clinical phantom studies, SmuSim effectively simulates distorted photon counting CT images. In digital physical phantom simulations, the deviations in attenuation coefficient due to charge sharing effects are −49.70%, −19.66%, and −3.33% for the three energy bins, respectively. In digital clinical phantom simulations, the differences in attenuation coefficient are −19.92%, −4.98%, and −0.6%, respectively. In the two simulation studies, the deviations between the results obtained from SmuSim and those from Monte Carlo simulation are less than 3% and 2%, respectively, demonstrating the effectiveness of the proposed SmuSim.
Conclusion
We analyze charge sharing effects in photon counting CT, a comprehensive analytical model, and finally simulate CT images with charge sharing effects for evaluation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Guangdong Provincial Key Laboratory of Medical Image Processing, School of Biomedical Engineering, Southern Medical University, Guangzhou, China.
- Pazhou Lab (Huangpu), Guangzhou, China.
- School of Integrated Circuits, Beijing University of Posts and Telecommunications, Beijing, China.
- Department of Radiation Oncology, Peking University Shenzhen Hospital, Shenzhen, China.
- Shenzhen Institute of Advanced Technology, Chinese Academy of Sciences, Shenzhen, China.
- School of Life Science and Technology, Xi'an Jiaotong University, Xi'an, China.
ORCID-Profile: Jiabing Sheng, Zhaoying Bian
Förderung
- Key Technologies Research and Development Program
- Guangzhou Municipal Science and Technology Project
- Science and Technology Planning Project of Shenzen Municipality
Zitieren
APA
Sheng, J., Zeng, D., Bian, Z., Li, M., Wu, Y., Li, X., Ge, Y., & Ma, J. (2025). Inclusion of spatio-energetic charge sharing effect model for accurate photon counting CT simulation. Journal of X-Ray Science and Technology. https://doi.org/10.1177/08953996251323725
Vancouver
Sheng J, Zeng D, Bian Z, Li M, Wu Y, Li X, et al. Inclusion of spatio-energetic charge sharing effect model for accurate photon counting CT simulation. Journal of X-Ray Science and Technology. 2025. doi:10.1177/08953996251323725
BibTeX
@article{Sheng2025,
author = {Sheng, Jiabing and Zeng, Dong and Bian, Zhaoying and Li, Mingqiang and Wu, Yongle and Li, Xin and Ge, YongShuai and Ma, Jianhua},
title = {Inclusion of spatio-energetic charge sharing effect model for accurate photon counting CT simulation},
journal = {Journal of X-Ray Science and Technology},
year = {2025},
doi = {10.1177/08953996251323725}
}
EUVIMED