EUVIMEDEUVIMED entdecken
Crossref

Unchecked Wall Vacuum Might Pose a Risk: An In Vitro Experimental Study on the Effects of Negative Wall Pressure on Suction Ureteral Access Sheaths—A PACE Group Study

Jorge Peña Lueza iD, Alejandra Bravo-Balado iD, Carlos Alonso Altez Fernández, Federico Zorzi iD, Aideen Madden, Luigi Candela, Marie Chicaud iD, Letizia Maria Ippolita Jannello, Stessy Kutchukian iD, Marie-Lou Letouche iD, Razvan Ionut Popescu, Alberto Quarà und 3 weitere

Journal of Endourology · 11.11.2026

DOI: 10.1177/08927790261505300 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Endourology, online erschienen am 11.11.2026. Verfasst von Jorge Peña Lueza und 14 weiteren Autor:innen. Beteiligte Einrichtung: Progress And Collaboration in Endourology (PACE) Group, Paris, France. Die Arbeit nennt 24 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

While safe positive intrarenal pressure (IRP) thresholds during retrograde intrarenal surgery (RIRS) are well known, negative outlet wall pressure applied to Flexible and Navigable Suction Sheaths (FANS) remains uncharacterized.

Objective

To evaluate IRP dynamics under varying irrigation heights and negative suction pressures, providing foundational data to support evidence-based protocols in suction-assisted endourology. Materials and Methods: A silicone kidney model was instrumented with an 11–13Fr/46 cm FANS and a 7.5Fr single-use flexible ureteroscope. Real-time IRP was measured via pressure wire during four sequential phases: irrigation (30 seconds), suction (30 seconds), simulated channel blockage (90 seconds), and unblocking (5 seconds). Four experimental groups evaluated combinations of different irrigation heights (40 and 100 cmH 2 O) and negative pressures (−100 and −375 mmHg). Each combination was replicated three times, yielding 12 independent experiments. The primary outcome was the rate of IRP change (mmHg/s) during unblocking, and the secondary outcome was the maximum IRP observed during blockage. Data were analyzed using one-way ANOVA with Tukey’s post hoc test ( p < 0.05).

Results

During unblocking, −375 mmHg suction produced significantly faster IRP reduction than −100 mmHg across all subgroups (mean differences 7.6–10.45 mmHg/s, p < 0.05). During blockage, 100 cmH 2 O irrigation generated significantly higher peak pressures than 40 cmH 2 O (mean differences 7.97–18.37 mmHg, p < 0.05). The combination of high irrigation pressure and high negative suction resulted in the most extreme and rapid pressure fluctuations, with changes occurring within fractions of 10-milliseconds.

Conclusions

Negative outlet wall pressure significantly influences IRP dynamics during RIRS with FANS, producing rapid and extreme pressure changes. Pressure changes may occur on a timescale that could challenge manual recognition and corrective intervention. Standardized pressure monitoring and automated control systems might become a promising approach to ensuring patient safety during suction-assisted endourology.

Schlagwörter

Stichwörter der Autor:innen

FANSRIRSSUASnegative pressurestonesuction

Autor:innen und Institutionen

  • Progress And Collaboration in Endourology (PACE) Group, Paris, France.
  • Faculty of Medicine (Alumnus), University of Lleida, Lleida, Spain.
  • Urology Department, Clínic Lleida, Lleida, Spain.
  • Department of Urology, Stone Unit, Fundació Puigvert, Autonomous University of Barcelona, Barcelona, Spain.
  • Servicio de Urología, Hospital Universitario A Coruña, La Coruña, Spain.
  • Department of Urology, University of Trieste, Trieste, Italy.

ORCID-Profile: Jorge Peña Lueza, Alejandra Bravo-Balado, Federico Zorzi, Marie Chicaud, Stessy Kutchukian, Marie-Lou Letouche, Mariela Corrales

Zitieren

APA

Peña Lueza, J., Bravo-Balado, A., Altez Fernández, C. A., Zorzi, F., Madden, A., Candela, L., Chicaud, M., Jannello, L. M. I., Kutchukian, S., Letouche, M. L., Popescu, R. I., Quarà, A., Corrales, M., Moretto, S., & Traxer, O. (2026). Unchecked Wall Vacuum Might Pose a Risk: An In Vitro Experimental Study on the Effects of Negative Wall Pressure on Suction Ureteral Access Sheaths—A PACE Group Study. Journal of Endourology. https://doi.org/10.1177/08927790261505300

Vancouver

Peña Lueza J, Bravo-Balado A, Altez Fernández CA, Zorzi F, Madden A, Candela L, et al. Unchecked Wall Vacuum Might Pose a Risk: An In Vitro Experimental Study on the Effects of Negative Wall Pressure on Suction Ureteral Access Sheaths—A PACE Group Study. Journal of Endourology. 2026. doi:10.1177/08927790261505300

BibTeX
@article{PeaLueza2026,
  author = {Peña Lueza, Jorge and Bravo-Balado, Alejandra and Altez Fernández, Carlos Alonso and Zorzi, Federico and Madden, Aideen and Candela, Luigi and Chicaud, Marie and Jannello, Letizia Maria Ippolita and Kutchukian, Stessy and Letouche, Marie-Lou and Popescu, Razvan Ionut and Quarà, Alberto and Corrales, Mariela and Moretto, Stefano and Traxer, Olivier},
  title = {Unchecked Wall Vacuum Might Pose a Risk: An In Vitro Experimental Study on the Effects of Negative Wall Pressure on Suction Ureteral Access Sheaths—A PACE Group Study},
  journal = {Journal of Endourology},
  year = {2026},
  doi = {10.1177/08927790261505300}
}
Kostenlos starten →