Early Evaluation of Safety and Efficacy of Robotic-Assisted Combined IntraRenal Surgery with the MONARCH TM Platform
Journal of Endourology · 2026
DOI: 10.1177/08927790261459280 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Endourology, 2026. Verfasst von Margaret A Knoedler und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Urology, University of Wisconsin School of Medicine and Public Health, Madison, Wisconsin, USA. Hauptthemen laut MEDLINE: Robotic Surgical Procedures, Nephrolithotomy, Percutaneous, Kidney Calculi, Kidney. Die Arbeit nennt 15 Literaturangaben. Gefördert von: Johnson and Johnson MedTech Surgery.
Abstract (Originalsprache)
Background and Objective: Percutaneous nephrolithotomy (PCNL) is the standard of care for treating complex renal stones, but only a minority of urologists in the United States independently obtain percutaneous access. The MONARCH™ Platform, Urology, is an FDA-cleared robotic platform that may simplify percutaneous access and efficiently treat complex renal stones. In this study, we evaluate the clinical, procedural, and patient-reported outcomes using this technology.
Methods
In a preliminary prospective study in two institutions (University of Wisconsin and Indiana University), patients were enrolled to undergo robotic-assisted mini-PCNL with combined ureteroscopic lithotripsy and followed up to 90 days post-procedure. Percutaneous access was gained using electromagnetic guidance while the patient was in a modified supine position. Stones were removed using a handheld controller to manipulate multiple instruments (i.e., retrograde ureteroscope, laser fiber, basket, flexible antegrade suction catheter) and control a fluid management system.
Results
Fifteen patients with median (range) linear stone burden of 32.6 (15.0–72.8) mm were enrolled. In 100% of cases, percutaneous access was successfully gained using this technology, and in 73.3% of cases this was achieved within a single attempt. Within this limited and highly heterogeneous study population, we achieved a postoperative day 30 median (range) volume reduction in stone burden of 98.8 (51.9–100.0)% and a stone-free rate (≤4 mm) of 50.0%. One perioperative Grade 3 complication occurred.
Conclusions
Our initial experience suggests a future where the MONARCH™ Platform may provide urologists with expanded capabilities for safe and effective treatment of select patients with nephrolithiasis.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Urology, University of Wisconsin School of Medicine and Public Health, Madison, Wisconsin, USA.
- Department of Urology, Indiana University School of Medicine, Indianapolis, Indiana, USA.
- Scientific Affairs, Johnson & Johnson MedTech Surgery, Santa Clara, California, USA.
ORCID-Profile: Margaret A Knoedler
Förderung
- Johnson and Johnson MedTech Surgery
Zitieren
APA
Knoedler, M. A., Serrell, E. C., Dalton, D. C., Sehgel, N. L., Morris, P. T., Schwiers, M. L., Antar, A., & Rivera, M. E. (2026). Early Evaluation of Safety and Efficacy of Robotic-Assisted Combined IntraRenal Surgery with the MONARCH TM Platform. Journal of Endourology. https://doi.org/10.1177/08927790261459280
Vancouver
Knoedler MA, Serrell EC, Dalton DC, Sehgel NL, Morris PT, Schwiers ML, et al. Early Evaluation of Safety and Efficacy of Robotic-Assisted Combined IntraRenal Surgery with the MONARCH TM Platform. Journal of Endourology. 2026. doi:10.1177/08927790261459280
BibTeX
@article{Knoedler2026,
author = {Knoedler, Margaret A and Serrell, Emily C and Dalton, David C and Sehgel, Nancy L and Morris, Paul T and Schwiers, Michael L and Antar, Ali and Rivera, Marcelino E},
title = {Early Evaluation of Safety and Efficacy of Robotic-Assisted Combined IntraRenal Surgery with the MONARCH TM Platform},
journal = {Journal of Endourology},
year = {2026},
doi = {10.1177/08927790261459280}
}
EUVIMED