Are Two Channels Better than One? A Comparative Study Between Dual-Lumen and Single-Lumen Flexible Ureteroscope in Retrograde Intrarenal Surgery

Dimitrios Diamantidis iD, Michael N Chrysikos iD, George Toufas, Chrysostomos Georgellis, Stavros Giannopoulos, Georgios Tsakaldimis, Christos Kalaitzis, Stilianos Giannakopoulos iD

Journal of Endourology · 01.08.2026 · 40 (8) · 873-878

DOI: 10.1177/08927790261451054 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Journal of Endourology, online erschienen am 21.05.2026. Verfasst von Dimitrios Diamantidis und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Urology, Democritus University of Thrace, Alexandroupolis, Greece. Hauptthemen laut MEDLINE: Ureteroscopes, Kidney Calculi, Ureteroscopy, Kidney. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 15 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

To compare perioperative efficiency, stone-free rates, and safety between a dual-lumen and a single-lumen flexible ureteroscope in retrograde intrarenal surgery (RIRS). Materials and Methods: This comparative study included 80 consecutive patients with a single renal stone 1.0–1.5 cm in the pelvicalyceal system treated with RIRS using either a dual-lumen ureteroscope (Group A, n = 40) or a single-lumen ureteroscope (Group B, n = 40). Recorded outcomes included operative and fluoroscopy times, failure of ureteral access sheath (UAS) insertion, surgeon-rated clarity of vision (0–10 scale), stone-free status on non-contrast computed tomography at 30 days, and complications graded by the Clavien–Dindo system and the Post-Ureteroscopy Lesion Scale (PULS). The stone-free rate (SFR) was defined as absolute and relative.

Results

Baseline demographics, stone size, and stone location were comparable between groups. Dual-lumen use was associated with shorter operative time (38.6 ± 9.3 vs 52.7 ± 8.1 minutes, p = 0.018), shorter fluoroscopy time (23.6 ± 5.5 vs 34.2 ± 4.7 seconds, p = 0.04), and higher clarity-of-vision scores (8.6 ± 1.35 vs 5.9 ± 1.86, p = 0.035). UAS insertion failed more often with the dual-lumen scope (20.0% vs 10.0%, p = 0.32). One month after evaluation, SFR did not significantly differ between the two groups. Postoperative complications were infrequent (12.5% vs 10.0%, p = 0.72), all were Clavien–Dindo grade I–II, and no PULS grade ≥4 ureteral injuries occurred.

Conclusions

Compared with a reusable single-lumen ureteroscope, the dual-lumen device improved operative efficiency in terms of shorter operative and fluoroscopy time and better clarity of vision. These benefits are counterbalanced by a higher UAS insertion failure rate related to the larger outer diameter of the ureteroscope.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

UreteroscopesKidney CalculiUreteroscopyKidneyHumansFemaleMaleMiddle AgedFluoroscopyAdult

Stichwörter der Autor:innen

dual-lumen ureteroscopeflexible ureteroscopykidney stone diseaseretrograde intrarenal surgerysingle-lumen ureteroscope

Autor:innen und Institutionen

  • Department of Urology, Democritus University of Thrace, Alexandroupolis, Greece.
  • 3rd Department of Urology, Interbalkan Medical Center, Thessaloniki, Greece.
  • Endoscopy Unit, Democritus University of Thrace, Alexandroupolis, Greece.

ORCID-Profile: Dimitrios Diamantidis, Michael N Chrysikos, Stilianos Giannakopoulos

Zitieren

APA

Diamantidis, D., Chrysikos, M. N., Toufas, G., Georgellis, C., Giannopoulos, S., Tsakaldimis, G., Kalaitzis, C., & Giannakopoulos, S. (2026). Are Two Channels Better than One? A Comparative Study Between Dual-Lumen and Single-Lumen Flexible Ureteroscope in Retrograde Intrarenal Surgery. Journal of Endourology, 40(8), 873-878. https://doi.org/10.1177/08927790261451054

Vancouver

Diamantidis D, Chrysikos MN, Toufas G, Georgellis C, Giannopoulos S, Tsakaldimis G, et al. Are Two Channels Better than One? A Comparative Study Between Dual-Lumen and Single-Lumen Flexible Ureteroscope in Retrograde Intrarenal Surgery. Journal of Endourology. 2026;40(8):873-878. doi:10.1177/08927790261451054

BibTeX
@article{Diamantidis2026,
  author = {Diamantidis, Dimitrios and Chrysikos, Michael N and Toufas, George and Georgellis, Chrysostomos and Giannopoulos, Stavros and Tsakaldimis, Georgios and Kalaitzis, Christos and Giannakopoulos, Stilianos},
  title = {Are Two Channels Better than One? A Comparative Study Between Dual-Lumen and Single-Lumen Flexible Ureteroscope in Retrograde Intrarenal Surgery},
  journal = {Journal of Endourology},
  year = {2026},
  volume = {40},
  number = {8},
  pages = {873--878},
  doi = {10.1177/08927790261451054}
}