Flexible Ureteroscopy for Renal Stones < 20 mm with Low-Power Holmium Laser: Is Basket Extraction still Needed or Is Dusting Enough? A Prospective Study with CT-Based Outcome Assessment
Journal of Endourology · 2025
DOI: 10.1177/08927790251390878 ↗Kurz gefasst
Randomisierte kontrollierte Studie in der Zeitschrift Journal of Endourology, 2025. Verfasst von Alex Meller und 5 weiteren Autor:innen. Beteiligte Einrichtung: Federal University of São Paulo (UNIFESP), São Paulo, Brazil. Hauptthemen laut MEDLINE: Ureteroscopy, Kidney Calculi, Lasers, Solid-State, Tomography, X-Ray Computed. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 29 Literaturangaben.
Abstract (Originalsprache)
Objective
This study compared the stone-free rate (SFR) and complications of two holmium laser lithotripsy techniques—active fragment extraction (FRAG) and dusting (DUST)—during flexible ureteroscopy (FURS) for renal stones using noncontrast computed tomography (NCCT). Materials and Methods: A total of 100 patients with renal stones (5–20 mm) undergoing FURS were randomized into FRAG and DUST groups. All surgical parameters, except lithotripsy strategy, were standardized, and ureteral access sheaths were used. The primary outcome was SFR at 90 days, defined as residual fragments ≤2 mm (Grade A + B) on NCCT. Secondary outcomes included the influence of stone volume, density, and complication rates.
Results
Ninety-three patients completed follow-up (FRAG: 45; DUST: 48). Baseline demographics and stone parameters (diameter, density, volume, location) were similar. Mean stone size (FRAG: 13.9 ± 5.2 mm vs DUST: 13.0 ± 4.7 mm, p = 0.378) and density (FRAG: 999.7 ± 267.6 Hounsfield unit [HU] vs DUST: 1,111.8 ± 249.1 HU, p = 0.044) were comparable. SFRs were 66.7% (FRAG) and 68.8% (DUST) ( p = 0.751). Multivariate analysis identified higher body mass index as a predictor of improved SFR (odds ratio 1.17, p = 0.022), whereas stone volume ( p = 0.179) and density ( p = 0.884) had no significant impact. Operative time, complications, and ancillary procedure rates were comparable.
Conclusions
Both techniques achieved similar SFRs, with no significant influence from stone volume or density. Postoperative complications and ancillary procedures were comparable. These findings suggest either method is effective, with technique selection based on surgeon preference.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Federal University of São Paulo (UNIFESP), São Paulo, Brazil.
- Tenon Hospital, Sorbonne University, Paris, France.
Zitieren
APA
Meller, A., Traxer, O., Freschi, G., Astolfi, R. H., Resuto, D., & Sadi, M. V. (2025). Flexible Ureteroscopy for Renal Stones < 20 mm with Low-Power Holmium Laser: Is Basket Extraction still Needed or Is Dusting Enough? A Prospective Study with CT-Based Outcome Assessment. Journal of Endourology. https://doi.org/10.1177/08927790251390878
Vancouver
Meller A, Traxer O, Freschi G, Astolfi RH, Resuto D, Sadi MV. Flexible Ureteroscopy for Renal Stones < 20 mm with Low-Power Holmium Laser: Is Basket Extraction still Needed or Is Dusting Enough? A Prospective Study with CT-Based Outcome Assessment. Journal of Endourology. 2025. doi:10.1177/08927790251390878
BibTeX
@article{Meller2025,
author = {Meller, Alex and Traxer, Olivier and Freschi, Gustavo and Astolfi, Rafael Haddad and Resuto, Diego and Sadi, Marcus V},
title = {Flexible Ureteroscopy for Renal Stones < 20 mm with Low-Power Holmium Laser: Is Basket Extraction still Needed or Is Dusting Enough? A Prospective Study with CT-Based Outcome Assessment},
journal = {Journal of Endourology},
year = {2025},
doi = {10.1177/08927790251390878}
}