Comparison of Ion Selective Electrodes for the Measurement of Sodium and Potassium in Human Milk
Journal of Human Lactation · 2026-05 · 42 (2) · 457-467
DOI: 10.1177/08903344261419335 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Journal of Human Lactation, online erschienen am 17.03.2026. Verfasst von Victoria Tennert und 6 weiteren Autor:innen. Beteiligte Einrichtung: University Medical Center Hamburg-Eppendorf, Hamburg, Germany. Hauptthemen laut MEDLINE: Milk, Human, Sodium, Potassium, Ion-Selective Electrodes. Die Arbeit nennt 29 Literaturangaben. Gefördert von: NICHD NIH HHS, USDA, National Institute of Food and Agriculture.
Abstract (Originalsprache)
Background
Accessible and cost-effective measures of human milk biomarkers are needed for lactation screening and management. Sodium (Na), potassium (K), and the sodium-potassium (Na/K) ratio in human milk show promise as indicators of secretory activation and mammary gland health. Research Aim: This validation study compares the accuracy and precision of handheld and benchtop ion selective electrodes (ISE) as compared to inductively coupled plasma mass spectrometry (ICP-MS) for the measurement of Na and K in defatted and whole milk.
Methods
Milk samples ( N = 27) were assessed for Na and K in defatted and whole milk using ICP-MS, handheld, and benchtop ISE instruments. Bland-Altman analysis was performed to compare Na, K and Na/K measurements in whole versus defatted milk and between instruments.
Results
All three assay methods provided high precision for Na and K concentrations in both whole and defatted milk. Na/K ratios were similar between whole and defatted milk for both ICP-MS and the handheld ISE. Measurements of Na and Na/K were systematically higher with both ISE instruments as compared to ICP-MS. However, the systematic difference was approximately three times greater in the benchtop ISE as compared to the handheld ISE. This resulted in identification of a larger number of cases of elevated mammary epithelium permeability (MEP) with the benchtop ISE as compared to handheld ISE or ICP-MS.
Conclusions
Meaningful assessment of Na and Na/K in whole milk is possible using the handheld ISE. Its portability and ease of use suggest potential for assessing Na and K in human milk at the point-of-care for use in both research and clinical applications.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- University Medical Center Hamburg-Eppendorf, Hamburg, Germany.
- University of Massachusetts Amherst, Amherst, MA, USA.
- Pioneer Valley Life Sciences Institute, Baystate Medical Center, Springfield, MA, USA.
- University of New Mexico Health Sciences Center, Albuquerque, USA.
ORCID-Profile: Victoria Tennert, Kathleen F Arcaro, Stephen J Eyles, Eva P Browne, Katie T Kivlighan
Förderung
- NICHD NIH HHS
- USDA
- National Institute of Food and Agriculture
- Eunice Kennedy Shriver National Institute of Child Health and Human Development
- Baden-Wuerttemberg Stiftung
- UMass Amherst Mass Spectrometry Core Facility
Zitieren
APA
Tennert, V., Arcaro, K. F., Eyles, S. J., Pentecost, B. T., Schneider, S. S., Browne, E. P., & Kivlighan, K. T. (2026). Comparison of Ion Selective Electrodes for the Measurement of Sodium and Potassium in Human Milk. Journal of Human Lactation, 42(2), 457-467. https://doi.org/10.1177/08903344261419335
Vancouver
Tennert V, Arcaro KF, Eyles SJ, Pentecost BT, Schneider SS, Browne EP, et al. Comparison of Ion Selective Electrodes for the Measurement of Sodium and Potassium in Human Milk. Journal of Human Lactation. 2026;42(2):457-467. doi:10.1177/08903344261419335
BibTeX
@article{Tennert2026,
author = {Tennert, Victoria and Arcaro, Kathleen F and Eyles, Stephen J and Pentecost, Brian T and Schneider, Sallie S and Browne, Eva P and Kivlighan, Katie T},
title = {Comparison of Ion Selective Electrodes for the Measurement of Sodium and Potassium in Human Milk},
journal = {Journal of Human Lactation},
year = {2026},
volume = {42},
number = {2},
pages = {457--467},
doi = {10.1177/08903344261419335}
}
EUVIMED