Distal Biceps Tendon Repair With Interlinked Knotless All-Suture Anchors Provides Greater Footprint Optimization and Higher Fixation Security Over Intramedullary Cortical Button Repair: A Biomechanical Study
The American Journal of Sports Medicine · 2025
DOI: 10.1177/03635465251365498 ↗Kurz gefasst
Fachartikel in der Zeitschrift The American Journal of Sports Medicine, 2025. Verfasst von Jim C Hsu und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedics & Rehabilitation, Yale School of Medicine, New Haven, Connecticut, USA. Hauptthemen laut MEDLINE: Suture Anchors, Tendon Injuries, Suture Techniques. Die Arbeit nennt 37 Literaturangaben.
Abstract (Originalsprache)
Background
Recent biomechanical investigations of distal biceps tendon repair (DBTR) constructs have typically evaluated time-zero fixation security, while tendon-bone repair footprint characterization has been limited. Consequently, interactions between repair construct design, fixation security, and repair footprint parameters remain minimally assessed. Purpose/Hypothesis: The purpose was to compare time-zero fixation security and repair footprint parameters between a new DBTR construct with 2 interlinked knotless all-suture anchors and an established DBTR construct with an intramedullary cortical button. It was hypothesized that the new interlinked twin-anchor repair technique would demonstrate greater time-zero fixation security and footprint optimization. Study Design: Controlled laboratory study.
Methods
A total of 20 cadaveric elbows in 2 matched groups underwent DBTR with either (1) twin interlinked knotless all-suture anchors or (2) a single intramedullary cortical button. Anatomic and repair footprints were digitally captured with a 3-dimensional coordinate-measuring machine. The repair constructs underwent cyclic loading and then were loaded to failure. Anatomic and repair footprint areas and their overlap, tendon-bone interface and total construct displacement, ultimate failure load, and failure mode were recorded. Anatomic footprint restoration and repair footprint accuracy were calculated.
Results
The interlinked knotless twin-anchor repair construct demonstrated a larger repair footprint area (55.1 ± 14.9 vs 35.2 ± 19.8 mm 2 , respectively; P = .032), greater anatomic footprint restoration (42.7% ± 12.9% vs 20.2% ± 9.4%, respectively; P = .003), lower tendon-bone interface displacement (3.2 ± 1.2 vs 12.4 ± 6.6 mm, respectively; P = .003), lower total construct displacement (5.5 ± 1.7 vs 13.9 ± 8.1 mm, respectively; P = .015), and higher ultimate failure load (468.3 ± 124.2 vs 313.2 ± 103.4 N, respectively; P = .001) compared with the single-button repair construct. The most common failure mode was knot slippage/suture breakage (60%) in the single-button group and suture-tendon interface failure (50%) in the twin-anchor group.
Conclusion
While this cadaveric study did not account for the effects of tendon-bone healing, the novel interlinked twin-anchor DBTR construct demonstrated greater time-zero fixation security, a larger repair footprint, and greater anatomic footprint restoration over the established single-button repair construct. Clinical Relevance: A DBTR construct with twin interlinked knotless all-suture anchors offers multiple features, including time-zero fixation security and footprint optimization, that may potentially improve clinical outcomes.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Orthopaedics & Rehabilitation, Yale School of Medicine, New Haven, Connecticut, USA.
- Orthopaedic Biomechanics Laboratory, Congress Medical Foundation, Pasadena, California, USA.
- Orthopaedic Biomechanics Laboratory, Department of Orthopaedic Surgery, Creighton University School of Medicine, Omaha, Nebraska, USA.
ORCID-Profile: Jim C Hsu, Michelle H McGarry
Zitieren
APA
Hsu, J. C., Fraipont, G. M., McGarry, M. H., Hung, V. T., Salandra, J. M., Adamson, G. J., & Lee, T. Q. (2025). Distal Biceps Tendon Repair With Interlinked Knotless All-Suture Anchors Provides Greater Footprint Optimization and Higher Fixation Security Over Intramedullary Cortical Button Repair: A Biomechanical Study. The American Journal of Sports Medicine. https://doi.org/10.1177/03635465251365498
Vancouver
Hsu JC, Fraipont GM, McGarry MH, Hung VT, Salandra JM, Adamson GJ, et al. Distal Biceps Tendon Repair With Interlinked Knotless All-Suture Anchors Provides Greater Footprint Optimization and Higher Fixation Security Over Intramedullary Cortical Button Repair: A Biomechanical Study. The American Journal of Sports Medicine. 2025. doi:10.1177/03635465251365498
BibTeX
@article{Hsu2025,
author = {Hsu, Jim C and Fraipont, Genevieve M and McGarry, Michelle H and Hung, Victor T and Salandra, Jonathan M and Adamson, Gregory J and Lee, Thay Q},
title = {Distal Biceps Tendon Repair With Interlinked Knotless All-Suture Anchors Provides Greater Footprint Optimization and Higher Fixation Security Over Intramedullary Cortical Button Repair: A Biomechanical Study},
journal = {The American Journal of Sports Medicine},
year = {2025},
doi = {10.1177/03635465251365498}
}
EUVIMED