CT angiography of the lower limbs: combining small field of view with large matrix size and iterative reconstruction further improves image quality of below-the-knee arteries

Michael Scheschenja iD, Alexander König, Simon Viniol, Moritz B Bastian iD, Joel Wessendorf, Andreas H Mahnken

Acta Radiologica · 2024

DOI: 10.1177/02841851241258655 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Acta Radiologica, 2024. Verfasst von Michael Scheschenja und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Diagnostic & Interventional Radiology, Philipps-University Marburg, Marburg, Germany. Hauptthemen laut MEDLINE: Computed Tomography Angiography, Lower Extremity, Radiographic Image Interpretation, Computer-Assisted. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 26 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Image quality and diagnostic accuracy in computed tomography angiography (CTA) reach their limits in imaging of below-the-knee vessels.

Purpose

To evaluate whether image quality in CTA of lower limbs is further improvable by combining side-separate reconstruction with a larger matrix size and whether resulting noise can be compromised with iterative reconstruction (IR). Material and Methods From CTA of the lower extremities of 26 patients (5 women, 21 men; mean age = 68.5 ± 10.3 years), the lower legs were reconstructed side-separately with different reconstruction algorithms and matrix sizes including filtered back projection (FBP) with a 512 × 512 matrix, FBP with a 1024 × 1024 matrix, IR (SAFIRE) with a 512 × 512 matrix, and IR (SAFIRE) with a 1024 × 1024 matrix. A total of 208 CT series were evaluated. Subjective image quality was assessed by two readers using a 5-point Likert scale. Image noise was assessed by measuring signal-to-noise and contrast-to-noise ratios.

Results

Subjective image quality was rated significantly higher when using a 1024 × 1024 matrix ( P < 0.001) and could further be increased with IR. Vessel sharpness was rated significantly better with a larger matrix ( P < 0.001). Visible and measured image noise was significantly higher with a 1024 × 1024 matrix but could be reduced by using IR ( P < 0.001), even to a level below FBP with a 512 × 512 matrix while reconstructing with a larger matrix ( P < 0.001).

Conclusion

Image quality, image noise, and vessel sharpness can be further improved in CTA of the lower extremities with side-separate reconstruction using a 1024 × 1024 matrix size and IR.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Computed Tomography AngiographyLower ExtremityRadiographic Image Interpretation, Computer-AssistedHumansFemaleMaleAgedMiddle AgedSignal-To-Noise RatioAlgorithmsAged, 80 and overContrast MediaImage Processing, Computer-Assisted

Stichwörter der Autor:innen

Computed tomography angiographyimage qualityiterative reconstructionlower extremitiesmatrix size

Substanzen

Contrast Media

Autor:innen und Institutionen

  • Department of Diagnostic & Interventional Radiology, Philipps-University Marburg, Marburg, Germany.

ORCID-Profile: Michael Scheschenja, Moritz B Bastian

Zitieren

APA

Scheschenja, M., König, A., Viniol, S., Bastian, M. B., Wessendorf, J., & Mahnken, A. H. (2024). CT angiography of the lower limbs: combining small field of view with large matrix size and iterative reconstruction further improves image quality of below-the-knee arteries. Acta Radiologica. https://doi.org/10.1177/02841851241258655

Vancouver

Scheschenja M, König A, Viniol S, Bastian MB, Wessendorf J, Mahnken AH. CT angiography of the lower limbs: combining small field of view with large matrix size and iterative reconstruction further improves image quality of below-the-knee arteries. Acta Radiologica. 2024. doi:10.1177/02841851241258655

BibTeX
@article{Scheschenja2024,
  author = {Scheschenja, Michael and König, Alexander and Viniol, Simon and Bastian, Moritz B and Wessendorf, Joel and Mahnken, Andreas H},
  title = {CT angiography of the lower limbs: combining small field of view with large matrix size and iterative reconstruction further improves image quality of below-the-knee arteries},
  journal = {Acta Radiologica},
  year = {2024},
  doi = {10.1177/02841851241258655}
}
Kostenlos starten →