Biventricular pseudoaneurysm after transcatheter aortic valve replacement
Asian Cardiovascular and Thoracic Annals · 2023
DOI: 10.1177/02184923231178717 ↗Kurz gefasst
Fachartikel in der Zeitschrift Asian Cardiovascular and Thoracic Annals, 2023. Verfasst von Hironobu Nishiori und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiovascular Surgery, Chiba University Hospital, Chiba, Japan. Hauptthemen laut MEDLINE: Transcatheter Aortic Valve Replacement, Aneurysm, False, Heart Valve Prosthesis Implantation, Aortic Valve Stenosis.
Abstract (Originalsprache)
A 90-year-old man with a horizontal aorta underwent trans-femoral transcatheter aortic valve replacement (TF-TAVR) using a 29-mm Evolut PRO (Medtronic, Dublin, Ireland) for aortic stenosis. The surgical aortic valve replacement was not selected due to the patient’s age and high frailty scale. The Evolut PRO was delivered over a stiff wire for TAVR placed in the left ventricle (LV). One month later, the patient presented with shortness of breath, and transthoracic echocardiography showed a biventricular pseudoaneurysm with the left to right shunt through the pseudoaneurysm (Figure 1A and 1B). Computed tomography showed a biventricular pseudoaneurysm (Figure 2). The patient underwent biventricular pseudoaneurysm repair (Figure 3A and 3B) and was discharged on postoperative day 64. The echocardiography six months after the surgery showed no residual shunt from the repaired pseudoaneurysm into the right ventricle (Figure 3C). LV pseudoaneurysms after TF-TAVR is rare. In the present case, the stiff wire for TAVR was pushed and pulled to adjust the prosthetic valve position at the horizontal aorta, which caused LV injury and led to biventricular pseudoaneurysm formation. Although the stiff wire features a spiral-shaped tip designed to prevent LV damage; however, caution is still warranted due to its rigidity. Currently, we use the cusp overlap technique to avoid excessive pushing and pulling of the wire, preventing LV injury.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Cardiovascular Surgery, Chiba University Hospital, Chiba, Japan.
ORCID-Profile: Hironobu Nishiori
Zitieren
APA
Nishiori, H., Matsuura, K., Yakita, Y., Kanda, T., & Matsumiya, G. (2023). Biventricular pseudoaneurysm after transcatheter aortic valve replacement. Asian Cardiovascular and Thoracic Annals. https://doi.org/10.1177/02184923231178717
Vancouver
Nishiori H, Matsuura K, Yakita Y, Kanda T, Matsumiya G. Biventricular pseudoaneurysm after transcatheter aortic valve replacement. Asian Cardiovascular and Thoracic Annals. 2023. doi:10.1177/02184923231178717
BibTeX
@article{Nishiori2023,
author = {Nishiori, Hironobu and Matsuura, Kaoru and Yakita, Yasunori and Kanda, Tomoyoshi and Matsumiya, Goro},
title = {Biventricular pseudoaneurysm after transcatheter aortic valve replacement},
journal = {Asian Cardiovascular and Thoracic Annals},
year = {2023},
doi = {10.1177/02184923231178717}
}
EUVIMED