Does Prophylactic Replacement of Voice Prosthesis Make Sense? A Study to Predict Prosthesis Lifetime
Otolaryngology–Head and Neck Surgery · 2023
DOI: 10.1177/01945998221116815 ↗Kurz gefasst
Fachartikel in der Zeitschrift Otolaryngology–Head and Neck Surgery, 2023. Verfasst von Anne N Heirman und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Head and Neck Oncology and Surgery, The Netherlands Cancer Institute, Amsterdam, the Netherlands. Hauptthemen laut MEDLINE: Larynx, Artificial. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 22 Literaturangaben.
Abstract (Originalsprache)
Objective
Voice prosthesis leakage significantly affects the quality of life of patients undergoing laryngectomy, causing insecurity and frequent unplanned hospital visits and costs. In this study, the concept of prophylactic voice prosthesis replacement was explored to prevent leakages. Study Design Retrospective cohort study.
Setting
Tertiary hospital.
Methods
This study included all patients who underwent laryngectomy between 2000 and 2012 in the Netherlands Cancer Institute. Device lifetimes and voice prosthesis replacements of a retrospective cohort were used to calculate the number of needed voice prostheses per patient per year to prevent 70% of the leakages by prophylactic replacement. Various strategies for the timing of prophylactic replacement were considered: adaptive strategies based on the individual patient's history of replacement and fixed strategies based on the results of patients with similar voice prosthesis or treatment characteristics.
Results
Patients used a median 3.4 voice prostheses per year (range, 0.1‐48.1). We found high inter‐ and intrapatient variability in device lifetime. When prophylactic replacement is applied, this would become a median 9.4 voice prostheses per year, which means replacement every 38 days, implying >6 additional voice prostheses per patient per year. The individual adaptive model showed that preventing 70% of the leakages was impossible for most patients and only a median 25% can be prevented. Monte‐Carlo simulations showed that prophylactic replacement is not feasible due to the high coefficient of variation (SD/mean) in device lifetime.
Conclusion
Based on our simulations, prophylactic replacement of voice prostheses is not feasible due to high inter‐ and intrapatient variation in device lifetime.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Head and Neck Oncology and Surgery, The Netherlands Cancer Institute, Amsterdam, the Netherlands.
- Department of Biometrics and Statistics, The Netherlands Cancer Institute, Amsterdam, the Netherlands.
- Amsterdam Center of Language and Communication, University of Amsterdam, Amsterdam, the Netherlands.
- Department of Otorhinolaryngology-Head and Neck Surgery, University of Groningen, University Medical Center Groningen, Groningen, the Netherlands.
- Department of Maxillofacial Surgery, Amsterdam University Medical Center, Amsterdam, the Netherlands.
Zitieren
APA
Heirman, A. N., van der Noort, V., van Son, R., Petersen, J. F., van der Molen, L., Halmos, G. B., Dirven, R., & van den Brekel, M. W. M. (2023). Does Prophylactic Replacement of Voice Prosthesis Make Sense? A Study to Predict Prosthesis Lifetime. Otolaryngology–Head and Neck Surgery. https://doi.org/10.1177/01945998221116815
Vancouver
Heirman AN, van der Noort V, van Son R, Petersen JF, van der Molen L, Halmos GB, et al. Does Prophylactic Replacement of Voice Prosthesis Make Sense? A Study to Predict Prosthesis Lifetime. Otolaryngology–Head and Neck Surgery. 2023. doi:10.1177/01945998221116815
BibTeX
@article{Heirman2023,
author = {Heirman, Anne N and van der Noort, Vincent and van Son, Rob and Petersen, Japke F and van der Molen, Lisette and Halmos, Gyorgy B and Dirven, Richard and van den Brekel, Michiel W M},
title = {Does Prophylactic Replacement of Voice Prosthesis Make Sense? A Study to Predict Prosthesis Lifetime},
journal = {Otolaryngology–Head and Neck Surgery},
year = {2023},
doi = {10.1177/01945998221116815}
}
EUVIMED