Monopolar Cautery Use in Pediatric Cochlear Implant Users
Otolaryngology–Head and Neck Surgery · 2023
DOI: 10.1177/01945998221108051 ↗Kurz gefasst
Systematische Übersicht in der Zeitschrift Otolaryngology–Head and Neck Surgery, 2023. Verfasst von J Cody Page und 4 weiteren Autor:innen. Beteiligte Einrichtung: Bobby R. Alford Department of Otolaryngology-Head and Neck Surgery, Baylor College of Medicine, Houston, Texas, USA. Hauptthemen laut MEDLINE: Cochlear Implants, Cochlear Implantation. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 27 Literaturangaben.
Abstract (Originalsprache)
Objectives
To determine the incidence and impact of monopolar cautery use in a cohort of pediatric cochlear implant (CI) users. Study Design Case series from a retrospective chart review and a systematic review of the literature.
Setting
Tertiary academic referral center.
Methods
CI patient charts from 2012 to 2021 were reviewed from a single pediatric hospital system to determine if monopolar cautery was used during a subsequent surgical procedure. In addition, a systematic review of the literature was performed to identify additional, relevant patients. Postoperative CI function was the primary outcome measure.
Results
In total, 190 patients underwent a surgical procedure following cochlear implantation in a single pediatric hospital system. Fifteen patients (7.9%) and 17 distinct surgical procedures were identified in which monopolar cautery was used. Seven of these 17 cases (41.2%) involved the head and neck, and 10 were performed below the clavicles. No patients experienced a device failure or a decline in CI performance following surgery. A systematic review identified an additional 4 patients who underwent a surgery that used monopolar cautery following cochlear implantation, and no change in CI function was identified.
Conclusions
The present study adds additional support to the notion that monopolar cautery does not necessarily injure CI functionality. While the most risk adverse strategy when planning a surgical procedure for a CI patient is to avoid monopolar cautery use altogether, the use of cautery should not immediately be associated with implant dysfunction.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Bobby R. Alford Department of Otolaryngology-Head and Neck Surgery, Baylor College of Medicine, Houston, Texas, USA.
- Baylor College of Medicine Medical School, Houston, Texas, USA.
- Division of Otolaryngology, Department of Surgery Texas Children's Hospital, Houston, Texas, USA.
- Department of Neurosurgery, Baylor College of Medicine, Houston, Texas, USA.
Zitieren
APA
Page, J. C., Chapel, A. C., Silva, R. C., Sullivan, J. C., & Sweeney, A. D. (2023). Monopolar Cautery Use in Pediatric Cochlear Implant Users. Otolaryngology–Head and Neck Surgery. https://doi.org/10.1177/01945998221108051
Vancouver
Page JC, Chapel AC, Silva RC, Sullivan JC, Sweeney AD. Monopolar Cautery Use in Pediatric Cochlear Implant Users. Otolaryngology–Head and Neck Surgery. 2023. doi:10.1177/01945998221108051
BibTeX
@article{Page2023,
author = {Page, J Cody and Chapel, A Claire and Silva, Rodrigo C and Sullivan, J Connor and Sweeney, Alex D},
title = {Monopolar Cautery Use in Pediatric Cochlear Implant Users},
journal = {Otolaryngology–Head and Neck Surgery},
year = {2023},
doi = {10.1177/01945998221108051}
}
EUVIMED