Methods and reagent-lot comparisons by regression analysis: False positives
Annals of Clinical Biochemistry: International Journal of Laboratory Medicine · 2026-11 · 63 (6) · 494-496
DOI: 10.1177/00045632261454285 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Annals of Clinical Biochemistry: International Journal of Laboratory Medicine, online erschienen am 26.05.2026. Verfasst von William A Sadler. Beteiligte Einrichtung: Nuclear Medicine Department, Christchurch Hospital, Christchurch, New Zealand. Hauptthemen laut MEDLINE: False Positive Reactions. Die Arbeit nennt 6 Literaturangaben.
Abstract (Originalsprache)
Background
The significance level specified for a statistical analysis indicates the expected false positive rate. Failure to confine false positives to their expected rates can lead to misleading conclusions and especially in studies investigating required sample sizes. A simulation study was used to investigate false positives associated with regression analysis.
Methods
Randomly generated X, Y pairs with zero underlying bias were subjected to Deming regression analysis. The design incorporated equal X, Y errors, unequal but parallel errors and non-parallel errors, each in combination with small (2:1), moderate (10:1) and large (2667:1) maximum:minimum range ratios. Bias (false positives in this context) was signified by failure of 95% confidence intervals (CIs) or, alternatively, the joint 95% slope, intercept confidence region (CR) to enclose target values.
Results
False positive rates assessed by CIs ranged from 6 to 10% and were clearly range ratio dependent, while those assessed by CRs were stable and very close to the expected 5% throughout.
Conclusions
CRs are distinctly more reliable than CIs in controlling false positive rates and should be the preferred option when using regression analysis in methods and reagent-lot comparisons. The computer program used to perform the study is freely available.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Nuclear Medicine Department, Christchurch Hospital, Christchurch, New Zealand.
ORCID-Profile: William A Sadler
Zitieren
APA
Sadler, W. A. (2026). Methods and reagent-lot comparisons by regression analysis: False positives. Annals of Clinical Biochemistry: International Journal of Laboratory Medicine, 63(6), 494-496. https://doi.org/10.1177/00045632261454285
Vancouver
Sadler WA. Methods and reagent-lot comparisons by regression analysis: False positives. Annals of Clinical Biochemistry: International Journal of Laboratory Medicine. 2026;63(6):494-496. doi:10.1177/00045632261454285
BibTeX
@article{Sadler2026,
author = {Sadler, William A},
title = {Methods and reagent-lot comparisons by regression analysis: False positives},
journal = {Annals of Clinical Biochemistry: International Journal of Laboratory Medicine},
year = {2026},
volume = {63},
number = {6},
pages = {494--496},
doi = {10.1177/00045632261454285}
}
EUVIMED