Multizentrische Studie in der Zeitschrift Annals of Clinical Biochemistry: International Journal of Laboratory Medicine, 2026. Verfasst von Qian Liu und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Laboratory Medicine, Lianyungang Clinical College, Jiangsu University & The Second People's Hospital of Lianyungang, Lianyungang, China. Hauptthemen laut MEDLINE: Lipids, Blood Chemical Analysis. Die Arbeit nennt 38 Literaturangaben. Gefördert von: the Research Project of the Lianyungang Health Commission, the Lianyungang Cancer Prevention and Treatment Technology Development Plan Project, the Aging Health Research Project of the Lianyungang Health Committee.
Abstract (Originalsprache)
Background
The six sigma model is widely used in laboratory quality management. For the first time, this study introduced total allowable error (TEa) from WS/T403-2024 and ‘desirable’ biological variation (BV) as dual quality goals to evaluate serum lipid analytes in six laboratories and develop individualized quality control (QC) strategies.
Methods
We collected internal quality control (IQC) and external quality assessment (EQA) data to calculate sigma values for each serum lipid analyte. Normalized sigma method decision charts were employed, and the Westgard sigma rule flow chart with batch size plus the quality goal index (QGI) guided individualized QC strategies and improvement plans.
Results
Under the same quality goal, different QC concentrations produced varying sigma values. Sigma values also differed significantly between the two quality goals. When WS/T403-2024 was applied, all analytes except triglycerides (TGs) showed lower sigma values than under ‘desirable’ BV. Normalized sigma method decision charts effectively highlighted these differences. Based on the Westgard sigma rule flow chart with batch size and QGI, individualized QC strategies were created, and priority improvement measures were proposed for analytes with sigma values below six.
Conclusions
The six sigma model is a valuable tool for laboratory quality management, guiding laboratories to enhance the detection capabilities of serum lipid analytes through targeted QC strategies and improvement measures.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Laboratory Medicine, Lianyungang Clinical College, Jiangsu University & The Second People's Hospital of Lianyungang, Lianyungang, China.
- Department of Laboratory Medicine, Lianyungang Clinical College, Xuzhou Medical University & The Second People's Hospital of Lianyungang, Lianyungang, China.
- Department of Laboratory Medicine, The Second People's Hospital of Lianyungang Affiliated with Kangda College of Nanjing Medical University, Lianyungang, China.
- Department of Laboratory Medicine, Lianyungang Clinical College, Bengbu Medical University & The Second People's Hospital of Lianyungang, Lianyungang, China.
- Department of Laboratory Medicine, Affiliated Lianyungang Clinical College of Nantong University, Lianyungang, China.
- Department of Laboratory Medicine, Nanping First Hospital, Nanping, China.
ORCID-Profile: Qian Liu, Fumeng Yang
Förderung
- the Research Project of the Lianyungang Health Commission
- the Lianyungang Cancer Prevention and Treatment Technology Development Plan Project
- the Aging Health Research Project of the Lianyungang Health Committee
- the Key Project of the Research and Development Fund of Kangda College of Nanjing Medical University
- the Lianyungang City Science and Technology Bureau’s Key Research and Development Project on Social Development
Zitieren
APA
Liu, Q., Lin, Y., Yang, F., Dai, Y., Hang, H., Wang, M., Hu, M., & Yang, F. (2026). Application of the six sigma model to evaluate the analytical performance of serum lipid analytes and design quality control strategies: A multi-centre study. Annals of Clinical Biochemistry: International Journal of Laboratory Medicine. https://doi.org/10.1177/00045632251350503
Vancouver
Liu Q, Lin Y, Yang F, Dai Y, Hang H, Wang M, et al. Application of the six sigma model to evaluate the analytical performance of serum lipid analytes and design quality control strategies: A multi-centre study. Annals of Clinical Biochemistry: International Journal of Laboratory Medicine. 2026. doi:10.1177/00045632251350503
BibTeX
@article{Liu2026,
author = {Liu, Qian and Lin, Yu and Yang, Fang and Dai, Yaping and Hang, Huan and Wang, Menglin and Hu, Ming and Yang, Fumeng},
title = {Application of the six sigma model to evaluate the analytical performance of serum lipid analytes and design quality control strategies: A multi-centre study},
journal = {Annals of Clinical Biochemistry: International Journal of Laboratory Medicine},
year = {2026},
doi = {10.1177/00045632251350503}
}
EUVIMED