Fachartikel in der Zeitschrift Annals of Clinical Biochemistry: International Journal of Laboratory Medicine, 2024. Verfasst von Leo Lam und 1 weiteren Autor:innen. Beteiligte Einrichtung: Chemical Pathology, LabPlus, Auckland City Hospital, Auckland, New Zealand. Hauptthemen laut MEDLINE: Troponin I, Immunoglobulins. Die Arbeit wurde bisher 20-mal zitiert (Crossref) und nennt 19 Literaturangaben.
Abstract (Originalsprache)
Introduction
Macrotroponin is increasingly recognised as a cause of confusion in interpreting high-sensitivity cardiac troponin (hs-cTnI) results. In this study, we sought to evaluate two practical approaches to detecting macrotroponin. These two approaches are PEG precipitation and SVM (support vector machine) analysis to classify discrepancies between hs-cTn assays.
Method
Residual serum and heparin plasma specimens (n = 483) with initially elevated hs-cTnI from hospital and community laboratories were retested on multiple hs-cTn platforms before and after PEG precipitation and Protein A immunoglobulin depletion. SVM analysis was conducted to identify a linear equation that best discriminated specimens with macrotroponin using a combination of results from two different hs-cTn assays.
Findings
The diagnostic performance of PEG precipitation was carried out using Protein A immunoglobulin depletion as the reference comparator. When a cutoff residual activity after PEG precipitation of ≤ 20% was used, this threshold carried a high specificity of 92% (confidence interval 83–98%; n = 189) using the Siemens hs-cTnI Vista assay and 95% specificity (86%–98%; n = 242) using the Abbott hs-cTnI Architect assay. SVM analysis generated a linear equation identifying macrotroponin specimens from results obtained on two hs-cTn assays. This approach can be highly specific, comparable to PEG precipitation when certain assay combinations and concentrations are used.
Conclusion
We describe and identify practical alternatives to detecting macrotroponin. These approaches can be optimised for high specificity, reducing the need for more complex laboratory methods.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Chemical Pathology, LabPlus, Auckland City Hospital, Auckland, New Zealand.
- Biochemistry, Middlemore Hospital Laboratories, Auckland, New Zealand.
- Biochemistry, Labtests, Auckland, New Zealand.
ORCID-Profile: Leo Lam
Zitieren
APA
Lam, L., & Kyle, C. (2024). Practical approaches to the detection of macrotroponin. Annals of Clinical Biochemistry: International Journal of Laboratory Medicine. https://doi.org/10.1177/00045632231197301
Vancouver
Lam L, Kyle C. Practical approaches to the detection of macrotroponin. Annals of Clinical Biochemistry: International Journal of Laboratory Medicine. 2024. doi:10.1177/00045632231197301
BibTeX
@article{Lam2024,
author = {Lam, Leo and Kyle, Campbell},
title = {Practical approaches to the detection of macrotroponin},
journal = {Annals of Clinical Biochemistry: International Journal of Laboratory Medicine},
year = {2024},
doi = {10.1177/00045632231197301}
}
EUVIMED