EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC 4.0

Development of Raman Calibration Model Without Culture Data for In-Line Analysis of Metabolites in Cell Culture Media

Risa Hara iD, Wataru Kobayashi, Hiroaki Yamanaka, Kodai Murayama iD, Soichiro Shimoda, Yukihiro Ozaki iD

Applied Spectroscopy · 2023

DOI: 10.1177/00037028231160197 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Applied Spectroscopy, 2023. Verfasst von Risa Hara und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Research and Development, Yokogawa Electric Corporation, Musashino, Japan. Hauptthemen laut MEDLINE: Cell Culture Techniques, Lactic Acid. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 44 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

In this study, we developed a method to build Raman calibration models without culture data for cell culture monitoring. First, Raman spectra were collected and then analyzed for the signals of all the mentioned analytes: glucose, lactate, glutamine, glutamate, ammonia, antibody, viable cells, media, and feed agent. Using these spectral data, the specific peak positions and intensities for each factor were detected. Next, according to the design of the experiment method, samples were prepared by mixing the above-mentioned factors. Raman spectra of these samples were collected and were used to build calibration models. Several combinations of spectral pretreatments and wavenumber regions were compared to optimize the calibration model for cell culture monitoring without culture data. The accuracy of the developed calibration model was evaluated by performing actual cell culture and fitting the in-line measured spectra to the developed calibration model. As a result, the calibration model achieved sufficiently good accuracy for the three components, glucose, lactate, and antibody (root mean square errors of prediction, or RMSEP = 0.23, 0.29, and 0.20 g/L, respectively). This study has presented innovative results in developing a culture monitoring method without using culture data, while using a basic conventional method of investigating the Raman spectra of each component in the culture media and then utilizing a design of experiment approach.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cell Culture TechniquesLactic AcidCalibrationGlucoseCulture MediaSpectrum Analysis, Raman

Stichwörter der Autor:innen

Bioprocess monitoringPATRaman spectroscopyin-line monitoringmodel optimizationprocess analytical technology

Substanzen

Lactic AcidGlucoseCulture Media

Autor:innen und Institutionen

  • Department of Research and Development, Yokogawa Electric Corporation, Musashino, Japan.
  • Department of Life Business, Yokogawa Electric Corporation, Musashino, Japan.
  • School of Biological and Environmental Sciences, Kwansei Gakuin University, Sanda, Japan.

ORCID-Profile: Risa Hara, Kodai Murayama, Yukihiro Ozaki

Zitieren

APA

Hara, R., Kobayashi, W., Yamanaka, H., Murayama, K., Shimoda, S., & Ozaki, Y. (2023). Development of Raman Calibration Model Without Culture Data for In-Line Analysis of Metabolites in Cell Culture Media. Applied Spectroscopy. https://doi.org/10.1177/00037028231160197

Vancouver

Hara R, Kobayashi W, Yamanaka H, Murayama K, Shimoda S, Ozaki Y. Development of Raman Calibration Model Without Culture Data for In-Line Analysis of Metabolites in Cell Culture Media. Applied Spectroscopy. 2023. doi:10.1177/00037028231160197

BibTeX
@article{Hara2023,
  author = {Hara, Risa and Kobayashi, Wataru and Yamanaka, Hiroaki and Murayama, Kodai and Shimoda, Soichiro and Ozaki, Yukihiro},
  title = {Development of Raman Calibration Model Without Culture Data for In-Line Analysis of Metabolites in Cell Culture Media},
  journal = {Applied Spectroscopy},
  year = {2023},
  doi = {10.1177/00037028231160197}
}
Kostenlos starten →