EUVIMEDEUVIMED entdecken
Crossref

In-Situ Elemental Composition Analysis of Large Inhalable Aerosol Using Laser Induced Breakdown Spectroscopy

James Sipich iD, Christian L'Orange, John Volckens, Azer Yalin

Applied Spectroscopy · 2023-03 · 77 (3) · 261-269

DOI: 10.1177/00037028221146804 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Applied Spectroscopy, online erschienen am 21.12.2022. Verfasst von James Sipich und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Mechanical Engineering, 3447Colorado State University, Fort Collins, CO, USA. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 38 Literaturangaben. Gefördert von: National Institute for Occupational Safety and Health.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The ability to obtain information on the composition of airborne particles is a necessary part of identifying and controlling risks from exposure to potentially toxic materials, especially in the workplace. However, very few aerosol sampling instruments can characterize elemental composition in real time or measure large inhalable particles with aerodynamic diameter exceeding 20 µm. Here, we present the development and validation of a method for real time elemental composition analysis of large inhalable particles using laser-induced breakdown spectroscopy (LIBS). The prototype sensor uses a passive inlet and an optical triggering system to ablate falling particles with an LIBS plasma. Particle composition is quantified based on collected emission spectra using a real-time material classification algorithm. The approach was validated with a set of 1480 experimental spectra from four different aerosol test materials. We have studied effects of varying detection thresholds and find operating conditions with good agreement to truth values (F 1 score ≥ 0.9). Details of the analysis method, including subtracting the spectral contribution from the air plasma and reasons for the infrequent misclassifications, are discussed. The LIBS elemental analysis can be combined with our previously demonstrated direct-reading particle sizer (DRPS) to provide a system capable of both counting, sizing, and elemental analysis of large inhalable particles.

Schlagwörter

Stichwörter der Autor:innen

AerosolLIBSbackground removallaser-induced breakdown spectroscopymaterial identification

Autor:innen und Institutionen

  • Department of Mechanical Engineering, 3447Colorado State University, Fort Collins, CO, USA.

ORCID-Profile: James Sipich

Förderung

  • National Institute for Occupational Safety and Health

Zitieren

APA

Sipich, J., L'Orange, C., Volckens, J., & Yalin, A. (2023). In-Situ Elemental Composition Analysis of Large Inhalable Aerosol Using Laser Induced Breakdown Spectroscopy. Applied Spectroscopy, 77(3), 261-269. https://doi.org/10.1177/00037028221146804

Vancouver

Sipich J, L'Orange C, Volckens J, Yalin A. In-Situ Elemental Composition Analysis of Large Inhalable Aerosol Using Laser Induced Breakdown Spectroscopy. Applied Spectroscopy. 2023;77(3):261-269. doi:10.1177/00037028221146804

BibTeX
@article{Sipich2023,
  author = {Sipich, James and L'Orange, Christian and Volckens, John and Yalin, Azer},
  title = {In-Situ Elemental Composition Analysis of Large Inhalable Aerosol Using Laser Induced Breakdown Spectroscopy},
  journal = {Applied Spectroscopy},
  year = {2023},
  volume = {77},
  number = {3},
  pages = {261--269},
  doi = {10.1177/00037028221146804}
}