EUVIMEDEUVIMED entdecken
Systematische Übersicht

A Systematic Review of Surgical Characteristics and Adverse Events of an Active, Transcutaneous Bone Conduction Device

Systematische Übersicht zu chirurgischen Merkmalen und unerwünschten Ereignissen eines aktiven, transkutanen Knochenleitungsgeräts

Alma Jukic iD, Christopher C Munhall iD, Shawn M Stevens iD

Annals of Otology, Rhinology & Laryngology · 2024-11 · 133 (11) · 956-966

DOI: 10.1177/00034894241283269 ↗

Auf einen Blick

Die systematische Übersicht zu Osia-Implantationen fand niedrige Komplikationsraten, relativ kurze Operationszeiten und gute audiometrische Ergebnisse. Unerwünschte Ereignisse traten in 20,1 % der Fälle auf.

Fragestellung

Welche frühen Ergebnisse sind mit der Osia-Implantation verbunden?

Methode

Systematische Übersicht nach PRISMA; 4 Datenbanken durchsucht: PubMed, Scopus, Cochrane CENTRAL und CINAHL.

Ergebnisse

18 Studien mit 336 Patient:innen. Mittleres Alter 37,9 Jahre. Operationszeit 71,6 Minuten; PTA-Gewinn 35,4 dB; funktioneller Gewinn bei hohen Frequenzen 16,1 dB; SRT-Verbesserung 19,1 dB.

Einschränkungen

Fortgeschrittene audiometrische Berichterstattung begrenzt; audiometrische Daten und PROMs heterogen berichtet.

Schlussfolgerung der Autor:innen

Die Osia-BCD war mit niedrigen Komplikationsraten, relativ kurzen Operationszeiten und guten audiometrischen und Sprachresultaten assoziiert, besonders mit hohem Frequenzgewinn >6 kHz.

Kernaussagen

  • Unerwünschte Ereignisse in 20,1 % der Fälle
  • Postoperative Infektionen: 5 %
  • Heterogene und begrenzte audiometrische Berichterstattung

Deutsche Zusammenfassung: KI-gestützt aus dem Original-Abstract erstellt; alle Zahlen wurden automatisch mit dem Original abgeglichen. Maßgeblich ist der Original-Abstract.

Kurz gefasst

Systematische Übersicht in der Zeitschrift Annals of Otology, Rhinology & Laryngology, online erschienen am 22.09.2024. Verfasst von Alma Jukic und 2 weiteren Autor:innen. Beteiligte Einrichtung: University of Arizona College of Medicine-Phoenix, Phoenix, AZ, USA. Hauptthemen laut MEDLINE: Bone Conduction. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 37 Literaturangaben.

Auf einen BlickAbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

A new, active transcutaneous bone conduction device (BCD) was FDA-approved in 2019 in the USA. This systematic review sought to evaluate early outcomes associated with Osia implantation.

Methods

A systematic review was conducted according to Preferred Reporting Items for Systematic Reviews and Meta-analyses (PRISMA) guidelines. Four databases were reviewed: PubMed, Scopus, Cochrane CENTRAL, and CINAHL. Studies were included if they described audiometric, surgical characteristics/complications, or adverse events associated with the Osia BCD. Exclusion criteria: non-English language studies, animal investigations, reviews/meta-analyses, case reports, database studies.

Results

Eighteen studies with 336 patients were included. Mean age at implantation was 37.9 years. About 79.5% of patients had MHL/CHL and 19.5% had SSD/SHL. Mean operative time was 71.6 minutes. Mean PTA gain from unaided conditions was 35.4 dB. Mean functional gain at high frequency (6 kHz and above) from aided conditions was 16.1 dB. Mean improvement in speech recognition thresholds was 19.1 dB from unaided conditions. Adverse events (all types) were reported in 20.1% of cases. Across all studies, the postoperative infection rate was 5%. About 2% of patients reported magnet retention issues. About 1.65% of cases were complicated by hematomas.

Conclusions

Under systematic literature review, the Osia BCD has been associated with low complication rates, relatively short operative times, and good audiometric and speech outcomes, notably high frequency gain >6 kHz. More advanced audiometric outcome reporting remains limited and audiometric data and patient reported outcome measures were reported heterogeneously.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Bone ConductionHumansHearing AidsPostoperative ComplicationsHearing Loss, ConductiveProsthesis Implantation

Stichwörter der Autor:innen

Osiaactive transcutaneous hearing devicebone conduction implantbone-anchored hearing aidconductive hearing losshearing loss

Autor:innen und Institutionen

  • University of Arizona College of Medicine-Phoenix, Phoenix, AZ, USA.
  • Department of Otolaryngology-Head and Neck Surgery, Medical University of South Carolina, Charleston, SC, USA.
  • Department of Otolaryngology and Skull Base Surgery-Barrow Neurological Institute, Phoenix, AZ, USA.

ORCID-Profile: Alma Jukic, Christopher C Munhall, Shawn M Stevens

Zitieren

APA

Jukic, A., Munhall, C. C., & Stevens, S. M. (2024). A Systematic Review of Surgical Characteristics and Adverse Events of an Active, Transcutaneous Bone Conduction Device. Annals of Otology, Rhinology & Laryngology, 133(11), 956-966. https://doi.org/10.1177/00034894241283269

Vancouver

Jukic A, Munhall CC, Stevens SM. A Systematic Review of Surgical Characteristics and Adverse Events of an Active, Transcutaneous Bone Conduction Device. Annals of Otology, Rhinology & Laryngology. 2024;133(11):956-966. doi:10.1177/00034894241283269

BibTeX
@article{Jukic2024,
  author = {Jukic, Alma and Munhall, Christopher C and Stevens, Shawn M},
  title = {A Systematic Review of Surgical Characteristics and Adverse Events of an Active, Transcutaneous Bone Conduction Device},
  journal = {Annals of Otology, Rhinology & Laryngology},
  year = {2024},
  volume = {133},
  number = {11},
  pages = {956--966},
  doi = {10.1177/00034894241283269}
}