Minimizing Thermal Damage During Thulium Laser-Assisted Partial Arytenoidectomy: Pulsed Versus Continuous Cutting in an Ex-Vivo Calf Model
Annals of Otology, Rhinology & Laryngology · 2023
DOI: 10.1177/00034894231157648 ↗Kurz gefasst
Fachartikel in der Zeitschrift Annals of Otology, Rhinology & Laryngology, 2023. Verfasst von Erica Barnett und 4 weiteren Autor:innen. Beteiligte Einrichtung: Harvard Medical School, Boston, MA, USA. Hauptthemen laut MEDLINE: Laser Therapy, Larynx. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 26 Literaturangaben.
Abstract (Originalsprache)
Objectives
The 2 µm-wavelength thulium laser is an effective cutter during partial arytenoidectomy, but thermal trauma can damage adjacent laryngeal tissue. Pulsing laser energy may reduce trauma when compared to continuous-wave cutting. This study measured temperature changes, thermal trauma, and time to complete partial arytenoidectomy, with and without pulsing, in an ex-vivo calf model.
Methods
Tissue temperature and time to complete a trans-cartilaginous cut were measured during partial arytenoidectomy on ex-vivo calf vocal folds (N = 24) using a thulium laser in continuous-wave (CW, N = 12) and pulsed-wave (PW, N = 12) modes. Energy was 5 W for CW and PW cuts; pulse-widths were 250, 500, and 750 ms. Thermal damage was analyzed histologically by measuring the depth of lactate dehydrogenase (LDH) inactivation perpendicular to the laser-cut edge at the vocal process. Paired t-tests compared CW and PW modes.
Results
Change in temperature was lower using CW (6.5°C) compared to PW modes (250 ms = 18°; 500 ms = 16°; 750 = 19°; P < .05). Trans-cartilaginous cuts were completed faster using CW (37 seconds) compared to PW (250 ms = 136 seconds; 500 ms = 61 seconds; 750 = 44 seconds; P < .05), and both modes delivered the same total Joules. The average depth of LDH depletion (thermal damage) was similar for all cuts.
Conclusions
1. Thulium laser cuts in continuous-mode unexpectedly produced less tissue heating yet created similar thermal damage than pulsed-mode cuts during simulated partial arytenoidectomy. 2. Trans-cartilaginous cuts were completed significantly faster in continuous-mode as compared to pulsed-mode cutting. 3. Pulsing the thulium laser does not minimize thermal damage compared to continuous wave cutting during thulium laser-assisted partial arytenoidectomy.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Harvard Medical School, Boston, MA, USA.
- Department of Surgery - Harvard Medical School, Center for Laryngeal Surgery and Voice Rehabilitation - Massachusetts General Hospital, Boston, MA, USA.
ORCID-Profile: James A Burns
Zitieren
APA
Barnett, E., Heaton, J. T., Petrillo, R. H., Purnell, P., & Burns, J. A. (2023). Minimizing Thermal Damage During Thulium Laser-Assisted Partial Arytenoidectomy: Pulsed Versus Continuous Cutting in an Ex-Vivo Calf Model. Annals of Otology, Rhinology & Laryngology. https://doi.org/10.1177/00034894231157648
Vancouver
Barnett E, Heaton JT, Petrillo RH, Purnell P, Burns JA. Minimizing Thermal Damage During Thulium Laser-Assisted Partial Arytenoidectomy: Pulsed Versus Continuous Cutting in an Ex-Vivo Calf Model. Annals of Otology, Rhinology & Laryngology. 2023. doi:10.1177/00034894231157648
BibTeX
@article{Barnett2023,
author = {Barnett, Erica and Heaton, James T and Petrillo, Robert H and Purnell, Philip and Burns, James A},
title = {Minimizing Thermal Damage During Thulium Laser-Assisted Partial Arytenoidectomy: Pulsed Versus Continuous Cutting in an Ex-Vivo Calf Model},
journal = {Annals of Otology, Rhinology & Laryngology},
year = {2023},
doi = {10.1177/00034894231157648}
}