Validation of Automated Software for Guttae Quantification in Fuchs Endothelial Corneal Dystrophy
Translational Vision Science & Technology · 2026
DOI: 10.1167/tvst.15.5.8 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift Translational Vision Science & Technology, 2026. Verfasst von Simran R Sarin und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ophthalmology and Visual Sciences, University of Iowa Carver College of Medicine, Iowa City, IA, USA. Hauptthemen laut MEDLINE: Fuchs' Endothelial Dystrophy, Software, Image Processing, Computer-Assisted, Cornea. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 44 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Purpose
Corneal guttae are the hallmark pathologic phenotype of Fuchs endothelial corneal dystrophy (FECD). Tools performing quantitative analysis of guttae and their changes over time are needed as medical therapies advance. We developed and validated a novel automated software platform that reports the guttae area ratio (GAR) in FECD specular microscopy images.
Methods
Patients with FECD with modified Krachmer grade (mKg) 2 to 5 disease and no prior keratoplasty were enrolled and imaged prospectively with one specular microscope unit. In a validation study, expert manual grading and automated software grading of GAR were compared for agreement. Additional images were analyzed to correlate automated GAR with disease grade and subclinical edema by Scheimpflug imaging.
Results
Nineteen images were included in the validation study. Mean manual GAR (0.400 ± 0.213) and mean automated GAR (0.395 ± 0.182) demonstrated equivalence (P = 0.003). Additionally, 446 images (109 eyes) demonstrated that automated GAR increases with worsening disease in mKg 2 to 5. Most mKg 4 images (n = 172, 93.5%; 49 eyes) and mKg 5 images (n = 59, 55.1%; 19 eyes) had no evidence of definite subclinical edema. Repeat image analysis demonstrated high repeatability of GAR (intraclass correlation = 0.85; 95% confidence interval, 0.78-0.90).
Conclusions
This study validates that automated software for guttae quantification in FECD specular microscopy images is equivalent to manual expert grading. GAR, an objective metric that increases in direct proportion to mKg, may precede development of subclinical edema. Translational Relevance: This software supports precise quantification of FECD disease secondary to guttae, allowing for objective measurement of disease progression and therapeutic efficacy.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Ophthalmology and Visual Sciences, University of Iowa Carver College of Medicine, Iowa City, IA, USA.
- ADCIS Evolucare Group, Saint-Contest, France.
- Iowa Lions Eye Bank, Coralville, IA, USA.
- Department of Biostatistics, College of Public Health, University of Iowa, Iowa City, IA, USA.
Zitieren
APA
Sarin, S. R., Abou Assi, J., Skeie, J. M., Deman, P., Zimmerman, M. B., McGranahan, D. M., Troyer, J. M., Schmidt, G. A., Potevin, F., & Greiner, M. A. (2026). Validation of Automated Software for Guttae Quantification in Fuchs Endothelial Corneal Dystrophy. Translational Vision Science & Technology. https://doi.org/10.1167/tvst.15.5.8
Vancouver
Sarin SR, Abou Assi J, Skeie JM, Deman P, Zimmerman MB, McGranahan DM, et al. Validation of Automated Software for Guttae Quantification in Fuchs Endothelial Corneal Dystrophy. Translational Vision Science & Technology. 2026. doi:10.1167/tvst.15.5.8
BibTeX
@article{Sarin2026,
author = {Sarin, Simran R and Abou Assi, Jérôme and Skeie, Jessica M and Deman, Pierre and Zimmerman, M Bridget and McGranahan, Dustin M and Troyer, Jody M and Schmidt, Gregory A and Potevin, François and Greiner, Mark A},
title = {Validation of Automated Software for Guttae Quantification in Fuchs Endothelial Corneal Dystrophy},
journal = {Translational Vision Science & Technology},
year = {2026},
doi = {10.1167/tvst.15.5.8}
}
EUVIMED