Comparative Analysis of Melbourne Rapid Fields Web-Browser Perimeter and Humphrey Field Analyzer in Chinese Glaucoma Patients
Translational Vision Science & Technology · 23.07.2025 · 14 (7) · 18
DOI: 10.1167/tvst.14.7.18 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Translational Vision Science & Technology, erschienen am 23.07.2025. Verfasst von De-Fu Chen und 9 weiteren Autor:innen. Beteiligte Einrichtung: National Clinical Research Center for Ocular Diseases, Eye Hospital, Wenzhou Medical University, Wenzhou, China. Hauptthemen laut MEDLINE: Glaucoma, Visual Field Tests, Visual Fields. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 50 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Purpose
This study aims to evaluate the efficacy and practicality of the Melbourne Rapid Fields (MRF) web-browser based perimeter software, compared to the conventional Humphrey Field Analyzer (HFA) 24-2 Swedish Interactive Threshold Algorithm (SITA) Standard Protocol in a hospital environment in China.
Methods
The study was conducted in a clinical practice setting from December 2023 to August 2024. A total of 113 Chinese speaking participants were enrolled, with 91 eyes from 91 patients included in the analysis. Participants underwent visual field (VF) testing using both HFA and MRF. MRF was conducted with computer audio guidance in Chinese. The primary outcome measures included mean deviation (MD), pattern standard deviation (PSD), test duration, and reliability indicators.
Results
MRF returned a significant correlation with HFA for MD (Intraclass Correlation Coefficients [ICC] = 0.91, 95% confidence interval [CI] = 0.82-0.95) and PSD (ICC = 0.88, 95% CI = 0.82-0.92). The average test duration was notably shorter for MRF (266 ± 41 seconds) compared to HFA (385 ± 83 seconds, P < 0.001). Bland-Altman analysis revealed a mean bias of -1.8 decibels (dB) for MD and 0.4 dB for PSD (HFA-MRF). Additionally, MRF demonstrated high sensitivity and specificity, achieving an area under the curve (AUC) of 0.92 and 0.93 for MD and PSD, respectively. Furthermore, MRF found fewer false positive results.
Conclusions
MRF offers a reliable, efficient, and patient-friendly alternative to traditional VF testing for Chinese patients. Its implementation could substantially improve access to glaucoma care, especially in underserved regions and for patients unable to attend regular clinic appointments. Translational Relevance: MRF enables accessible, efficient glaucoma care, offering a reliable alternative to HFA, particularly for underserved Chinese populations.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- National Clinical Research Center for Ocular Diseases, Eye Hospital, Wenzhou Medical University, Wenzhou, China.
- Glaucoma Research Institute, Wenzhou Medical University, Wenzhou, Zhejiang, China.
- Royal Victorian Eye and Ear Hospital, East Melbourne, Victoria, Australia.
- Department of Optometry and Vision Science, University of Melbourne, Parkville, Victoria, Australia.
Zitieren
APA
Chen, D. F., Zhou, W., Lu, X., Wang, T., Wang, C., Wang, Y., Kong, Y. X. G., Vingrys, A. J., Ye, C., & Liang, Y. (2025). Comparative Analysis of Melbourne Rapid Fields Web-Browser Perimeter and Humphrey Field Analyzer in Chinese Glaucoma Patients. Translational Vision Science & Technology, 14(7), 18. https://doi.org/10.1167/tvst.14.7.18
Vancouver
Chen DF, Zhou W, Lu X, Wang T, Wang C, Wang Y, et al. Comparative Analysis of Melbourne Rapid Fields Web-Browser Perimeter and Humphrey Field Analyzer in Chinese Glaucoma Patients. Translational Vision Science & Technology. 2025;14(7):18. doi:10.1167/tvst.14.7.18
BibTeX
@article{Chen2025,
author = {Chen, De-Fu and Zhou, Wenzhe and Lu, Xiaonan and Wang, Tao and Wang, Chenmin and Wang, Yaoyao and Kong, Yu Xiang George and Vingrys, Algis J and Ye, Cong and Liang, Yuanbo},
title = {Comparative Analysis of Melbourne Rapid Fields Web-Browser Perimeter and Humphrey Field Analyzer in Chinese Glaucoma Patients},
journal = {Translational Vision Science & Technology},
year = {2025},
volume = {14},
number = {7},
pages = {18},
doi = {10.1167/tvst.14.7.18}
}
EUVIMED