Ocular Blood Velocity Measurement With Optical Coherence Tomography Using Speckle Analysis
Translational Vision Science & Technology · 2025
DOI: 10.1167/tvst.14.11.13 ↗Kurz gefasst
Fachartikel in der Zeitschrift Translational Vision Science & Technology, 2025. Verfasst von Richard F Spaide und 7 weiteren Autor:innen. Beteiligte Einrichtung: Vitreous Retina Macula Consultants of New York, New York, NY, USA. Hauptthemen laut MEDLINE: Tomography, Optical Coherence, Retinal Vessels, Eye, Retinal Vein Occlusion. Die Arbeit nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Purpose
To develop and validate a novel method for quantifying blood flow velocity in ocular vessels using speckle pattern analysis in optical coherence tomography (OCT) images.
Methods
We developed an OCT-based methodology that quantifies flow velocity by analyzing speckle structures within OCT images to derive a relative flow value (RFV) based on the horizontal derivative of speckle patterns. A model eye with 150-µm capillary tubing was used to evaluate speckle pattern analysis at controlled flow rates of human and porcine blood. We assessed the correlation between RFV and actual flow velocities and examined measurement consistency across vessel sectioning angles varying ±20° from perpendicular.
Results
RFV exhibited a power-law relationship with actual flow velocities, RFV = 169.6 × (Actual Flow)0.264 (r2 = 0.983), and maintained accuracy without saturation up to 70 mm/s. Measurements remained consistent across vessel sectioning angles. Intraluminal speckle structures changed dynamically with flow velocity, and averaged images revealed characteristic hourglass profiles at higher velocities. In a patient with central retinal vein occlusion, RFV detected altered pulsatility and reduced venous flow, aligning with clinical expectations.
Conclusions
This speckle-based OCT method provides a non-invasive, depth-resolved approach for quantifying ocular blood velocity with minimal computational requirements. It holds potential for diagnosing and monitoring vascular abnormalities in ocular and systemic diseases. Further validation is necessary across diverse patient populations. Translational Relevance: This method simplifies blood flow quantification, enhancing the role of OCT in hemodynamic assessment both ocularly and systemically.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Vitreous Retina Macula Consultants of New York, New York, NY, USA.
- Topcon Healthcare, Oakland, NJ, USA.
- Topcon Corporation, Tokyo, Japan.
Zitieren
APA
Spaide, R. F., Lee, K., Kuo, J. W., Akiba, M., Durbin, M., Naderi, A., Ko, T. H., & Tafreshi, A. (2025). Ocular Blood Velocity Measurement With Optical Coherence Tomography Using Speckle Analysis. Translational Vision Science & Technology. https://doi.org/10.1167/tvst.14.11.13
Vancouver
Spaide RF, Lee K, Kuo JW, Akiba M, Durbin M, Naderi A, et al. Ocular Blood Velocity Measurement With Optical Coherence Tomography Using Speckle Analysis. Translational Vision Science & Technology. 2025. doi:10.1167/tvst.14.11.13
BibTeX
@article{Spaide2025,
author = {Spaide, Richard F and Lee, Kyungmoo and Kuo, Jen-Wei and Akiba, Masahiro and Durbin, Mary and Naderi, Amirreza and Ko, Tony H and Tafreshi, Ali},
title = {Ocular Blood Velocity Measurement With Optical Coherence Tomography Using Speckle Analysis},
journal = {Translational Vision Science & Technology},
year = {2025},
doi = {10.1167/tvst.14.11.13}
}
EUVIMED