High Prevalence of Artifacts in Optical Coherence Tomography With Adequate Signal Strength
Translational Vision Science & Technology · 2024
DOI: 10.1167/tvst.13.8.43 ↗Kurz gefasst
Fachartikel in der Zeitschrift Translational Vision Science & Technology, 2024. Verfasst von Wei-Chun Lin und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ophthalmology, Casey Eye Institute, Oregon Health & Science University, Portland, OR, USA. Hauptthemen laut MEDLINE: Tomography, Optical Coherence, Artifacts, Deep Learning. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: NIH HHS, NEI NIH HHS.
Abstract (Originalsprache)
Purpose
This study aims to investigate the prevalence of artifacts in optical coherence tomography (OCT) images with acceptable signal strength and evaluate the performance of supervised deep learning models in improving OCT image quality assessment.
Methods
We conducted a retrospective study on 4555 OCT images from 546 patients, with each image having an acceptable signal strength (≥6). A comprehensive analysis of prevalent OCT artifacts was performed, and five pretrained convolutional neural network models were trained and tested to infer images based on quality.
Results
Our results showed a high prevalence of artifacts in OCT images with acceptable signal strength. Approximately 21% of images were labeled as nonacceptable quality. The EfficientNetV2 model demonstrated superior performance in classifying OCT image quality, achieving an area under the receiver operating characteristic curve of 0.950 ± 0.007 and an area under the precision recall curve of 0.985 ± 0.002.
Conclusions
The findings highlight the limitations of relying solely on signal strength for OCT image quality assessment and the potential of deep learning models in accurately classifying image quality. Translational Relevance: Application of the deep learning-based OCT image quality assessment models may improve the OCT image data quality for both clinical applications and research.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Ophthalmology, Casey Eye Institute, Oregon Health & Science University, Portland, OR, USA.
- Department of Pharmacology and Neuroscience, North Texas Eye Research Institute, University of North Texas Health Science Center, Fort Worth, TX, USA.
- Wills Eye Hospital, Philadelphia, PA, USA.
- Sidney Kimmel Medical College at Thomas Jefferson University, Philadelphia, PA, USA.
- Drexel University School of Biomedical Engineering, Sciences and Health Systems, Philadelphia, PA, USA.
- Department of Medical Informatics and Clinical Epidemiology (DMICE), Oregon Health & Science University, Portland, OR, USA.
Förderung
- NIH HHS
- NEI NIH HHS
Zitieren
APA
Lin, W. C., Coyner, A. S., Amankwa, C. E., Lucero, A., Wollstein, G., Schuman, J. S., & Ishikawa, H. (2024). High Prevalence of Artifacts in Optical Coherence Tomography With Adequate Signal Strength. Translational Vision Science & Technology. https://doi.org/10.1167/tvst.13.8.43
Vancouver
Lin WC, Coyner AS, Amankwa CE, Lucero A, Wollstein G, Schuman JS, et al. High Prevalence of Artifacts in Optical Coherence Tomography With Adequate Signal Strength. Translational Vision Science & Technology. 2024. doi:10.1167/tvst.13.8.43
BibTeX
@article{Lin2024,
author = {Lin, Wei-Chun and Coyner, Aaron S and Amankwa, Charles E and Lucero, Abigail and Wollstein, Gadi and Schuman, Joel S and Ishikawa, Hiroshi},
title = {High Prevalence of Artifacts in Optical Coherence Tomography With Adequate Signal Strength},
journal = {Translational Vision Science & Technology},
year = {2024},
doi = {10.1167/tvst.13.8.43}
}
EUVIMED