Fachartikel in der Zeitschrift Translational Vision Science & Technology, 2024. Verfasst von Iniya Adhan und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ophthalmology & Visual Sciences, Medical College of Wisconsin, Milwaukee, WI, USA. Hauptthemen laut MEDLINE: Retinal Cone Photoreceptor Cells, Fovea Centralis. Die Arbeit nennt 48 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NCRR NIH HHS, NEI NIH HHS.
Abstract (Originalsprache)
Purpose
To assess longitudinal reproducibility of metrics of foveal density (peak cone density [PCD], cone density centroid [CDC], and 80th percentile centroid area) in participants with normal vision.
Methods
Participants (n = 19; five male and 14 female) were imaged at two time points (average interval of 3.2 years) using an adaptive optics scanning light ophthalmoscope (AOSLO). Foveally centered regions of interest (ROIs) were extracted from AOSLO montages. Cone coordinate matrices were semiautomatically derived for each ROI, and cone mosaic metrics were calculated.
Results
On average, there were no significant changes in cone mosaic metrics between visits. The average ± SD PCD was 187,000 ± 20,000 cones/mm2 and 189,000 ± 21,700 cones/mm2 for visits 1 and 2, respectively (P = 0.52). The average ± SD density at the CDC was 183,000 ± 19,000 cones/mm2 and 184,000 ± 20,800 cones/mm2 for visits 1 and 2, respectively (P = 0.78). The average ± SD 80th percentile isodensity contour area was 15,400 ± 1800 µm2 and 15,600 ± 1910 µm2 for visits 1 and 2, respectively (P = 0.57).
Conclusions
Foveal cone mosaic density metrics were highly reproducible in the cohort examined here, although further study is required in more diverse populations. Translational Relevance: Determination of the normative longitudinal changes in foveal cone topography is key for evaluating longitudinal measures of foveal cone topography in patients with progressive retinal dystrophies.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Ophthalmology & Visual Sciences, Medical College of Wisconsin, Milwaukee, WI, USA.
- Joint Department of Biomedical Engineering, Marquette University and Medical College of Wisconsin, Milwaukee, WI, USA.
- Department of Cell Biology, Neurobiology & Anatomy, Medical College of Wisconsin, Milwaukee, WI, USA.
Förderung
- NCRR NIH HHS
- NEI NIH HHS
Zitieren
APA
Adhan, I., Warr, E., Grieshop, J., Kreis, J., Nikezic, D., Walesa, A., Hemsworth, K., Cooper, R. F., & Carroll, J. (2024). Intervisit Reproducibility of Foveal Cone Density Metrics. Translational Vision Science & Technology. https://doi.org/10.1167/tvst.13.6.18
Vancouver
Adhan I, Warr E, Grieshop J, Kreis J, Nikezic D, Walesa A, et al. Intervisit Reproducibility of Foveal Cone Density Metrics. Translational Vision Science & Technology. 2024. doi:10.1167/tvst.13.6.18
BibTeX
@article{Adhan2024,
author = {Adhan, Iniya and Warr, Emma and Grieshop, Jenna and Kreis, Joseph and Nikezic, Danica and Walesa, Ashleigh and Hemsworth, Katherine and Cooper, Robert F and Carroll, Joseph},
title = {Intervisit Reproducibility of Foveal Cone Density Metrics},
journal = {Translational Vision Science & Technology},
year = {2024},
doi = {10.1167/tvst.13.6.18}
}
EUVIMED