Acoustic Radiation Force Optical Coherence Elastography of the Crystalline Lens: Safety
Translational Vision Science & Technology · 2024
DOI: 10.1167/tvst.13.12.36 ↗Kurz gefasst
Fachartikel in der Zeitschrift Translational Vision Science & Technology, 2024. Verfasst von Christian Zevallos-Delgado und 11 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, University of Houston, Houston, TX, USA. Hauptthemen laut MEDLINE: Elasticity Imaging Techniques, Tomography, Optical Coherence, Lens, Crystalline, Microscopy, Confocal. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 39 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: NEI NIH HHS.
Abstract (Originalsprache)
Purpose
To assess the safety of acoustic radiation force optical coherence elastography in the crystalline lens in situ.
Methods
Acoustic radiation force (ARF) produced by an immersion single-element ultrasound transducer (nominal frequency = 3.5 MHz) was characterized using a needle hydrophone and used for optical coherence elastography (OCE) of the crystalline lens. Preamplified signals at 50, 100, 250, 500, 750, 1000, and 1250 mV peak amplitude were tested on ex vivo porcine eyes (n = 21). Three-dimensional optical coherence tomography (OCT) and confocal microscopy images were acquired before and after ARF exposure to each signal amplitude to determine damage.
Results
The acoustic intensity of the ultrasound transducer at 100-mV preamplified peak amplitude input demonstrated a signal-to-noise ratio high enough for tracking elastic wave propagation in the lens and spatial-peak pulse-average (SPPA) intensity of 24.1 W/cm² and mechanical index (MI) of 0.46. The SPPA intensity was lower than the U.S. Food and Drug Administration (FDA) safety limit (28 W/cm2), but the MI was twice the safety limit (0.23). OCT structural and confocal microscopy images showed damage only at levels exceeding 1150 W/cm2 and 3.2 for SPPA intensity and MI, respectively.
Conclusions
OCT and confocal microscopy showed that, even when the intensity exceeded FDA recommendations (>100 mV), no noticeable damage was observed. Although a further reduction in acoustic intensity is necessary to meet FDA safety limits, ARF-based elastography shows promise for safe clinical translation in quantitatively characterizing lenticular biomechanical properties. Translational Relevance: This work assessed the safety standards for acoustic radiation force to be used in human lens elastography according to the FDA safety limits.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Biomedical Engineering, University of Houston, Houston, TX, USA.
- College of Optometry, University of Houston, Houston, TX, USA.
- Ophtalmic Biophysics Center, Bascom Palmer Eye Institute, University of Miami Miller School of Medicine, Miami, FL, USA.
- Department of Biomedical Engineering, University of Miami College of Engineering, Miami, FL, USA.
- Fischell Department of Bioengineering Brain and Behavior Institute, University of Maryland, College Park, MD, USA.
- Department of Mechanical Engineering, University of Houston, Houston, TX, USA.
Förderung
- NEI NIH HHS
Zitieren
APA
Zevallos-Delgado, C., Mekonnen, T. T., Duvvuri, C., Rohman, L., Schumacher, J., Singh, M., Aglyamov, S. R., Twa, M. D., Parel, J. M., Scarcelli, G., Manns, F., & Larin, K. V. (2024). Acoustic Radiation Force Optical Coherence Elastography of the Crystalline Lens: Safety. Translational Vision Science & Technology. https://doi.org/10.1167/tvst.13.12.36
Vancouver
Zevallos-Delgado C, Mekonnen TT, Duvvuri C, Rohman L, Schumacher J, Singh M, et al. Acoustic Radiation Force Optical Coherence Elastography of the Crystalline Lens: Safety. Translational Vision Science & Technology. 2024. doi:10.1167/tvst.13.12.36
BibTeX
@article{ZevallosDelgado2024,
author = {Zevallos-Delgado, Christian and Mekonnen, Taye Tolu and Duvvuri, Chaitanya and Rohman, Leana and Schumacher, Justin and Singh, Manmohan and Aglyamov, Salavat R and Twa, Michael D and Parel, Jean-Marie and Scarcelli, Giuliano and Manns, Fabrice and Larin, Kirill V},
title = {Acoustic Radiation Force Optical Coherence Elastography of the Crystalline Lens: Safety},
journal = {Translational Vision Science & Technology},
year = {2024},
doi = {10.1167/tvst.13.12.36}
}
EUVIMED