EUVIMEDEUVIMED entdecken
Multizentrische StudieVergleichsstudieOpen Access · CC BY-NC-ND 4.0

Objective Visual Acuity Estimates in Amblyopia Are More Accurate With Optotype-Based P300 Than With VEP Measurements

Akshara V Gopiswaminathan, Julia Haldina, Khaldoon O Al-Nosairy, Céline Z Duval, Francie H Stolle, Michael B Hoffmann, Sven P Heinrich

Translational Vision Science & Technology · 2024

DOI: 10.1167/tvst.13.12.30 ↗

Kurz gefasst

Multizentrische Studie in der Zeitschrift Translational Vision Science & Technology, 2024. Verfasst von Akshara V Gopiswaminathan und 6 weiteren Autor:innen. Beteiligte Einrichtung: Section for Clinical and Experimental Sensory Physiology, Department of Ophthalmology, Otto-von-Guericke University, Magdeburg, Germany. Hauptthemen laut MEDLINE: Amblyopia, Visual Acuity, Evoked Potentials, Visual, Event-Related Potentials, P300. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Traditional visual acuity (VA) measurements depend on subjective responses, which can be unreliable, especially with uncooperative participants. Objective measurements with visual evoked potentials (VEP) address this issue but can overestimate VA in amblyopia. This study aims to establish the P300 component of the event-related potential as an objective VA test for amblyopia and compare its performance to subjective (psychophysical) and VEP-based VA estimates.

Methods

Psychophysical, VEP-based, and P300-based VA estimates were obtained for amblyopic and fellow eyes of 18 participants (aged 19-65) in a bicentric study. VEP-based VA was determined from the spatial frequency threshold derived from occipital cortex pattern-pulse responses to check-sizes ranging from 0.048° to 8.95°. P300 responses were collected using visual oddball sequences with circular optotypes. The threshold was estimated from the sigmoid function of parietal P300 amplitude versus optotype gap size. Mean VA values for amblyopic eyes were compared across methods.

Results

VEP-based VA of the amblyopic eyes overestimated psychophysical VA by 0.18 ± 0.06 logMAR (P = 0.0016). In contrast, P300-based VA showed no significant difference from psychophysical VA (0.00 ± 0.04 logMAR, P > 0.05).

Conclusions

In amblyopia, P300-based optotype VA aligns more closely with psychophysical VA than VEP-based VA, suggesting that P300-based VA is a valid objective alternative for estimating VA in amblyopic eyes. Translational Relevance: This study highlights the potential of P300-based VA testing as a reliable and objective method for assessing VA in amblyopic eyes, offering a promising tool for clinical and research applications where traditional methods fall short.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

AmblyopiaVisual AcuityEvoked Potentials, VisualEvent-Related Potentials, P300HumansFemaleMaleAdultYoung AdultMiddle AgedAgedVision Tests

Autor:innen und Institutionen

  • Section for Clinical and Experimental Sensory Physiology, Department of Ophthalmology, Otto-von-Guericke University, Magdeburg, Germany.
  • Eye Center, Medical Center, University of Freiburg, Freiburg, Germany.
  • Faculty of Medicine, University of Freiburg, Freiburg, Germany.
  • Center for Behavioral Brain Sciences, Magdeburg, Germany.

Zitieren

APA

Gopiswaminathan, A. V., Haldina, J., Al-Nosairy, K. O., Duval, C. Z., Stolle, F. H., Hoffmann, M. B., & Heinrich, S. P. (2024). Objective Visual Acuity Estimates in Amblyopia Are More Accurate With Optotype-Based P300 Than With VEP Measurements. Translational Vision Science & Technology. https://doi.org/10.1167/tvst.13.12.30

Vancouver

Gopiswaminathan AV, Haldina J, Al-Nosairy KO, Duval CZ, Stolle FH, Hoffmann MB, et al. Objective Visual Acuity Estimates in Amblyopia Are More Accurate With Optotype-Based P300 Than With VEP Measurements. Translational Vision Science & Technology. 2024. doi:10.1167/tvst.13.12.30

BibTeX
@article{Gopiswaminathan2024,
  author = {Gopiswaminathan, Akshara V and Haldina, Julia and Al-Nosairy, Khaldoon O and Duval, Céline Z and Stolle, Francie H and Hoffmann, Michael B and Heinrich, Sven P},
  title = {Objective Visual Acuity Estimates in Amblyopia Are More Accurate With Optotype-Based P300 Than With VEP Measurements},
  journal = {Translational Vision Science & Technology},
  year = {2024},
  doi = {10.1167/tvst.13.12.30}
}
Kostenlos starten →