EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC-ND 4.0

Impact of the Minimization of Standard Deviation Before Zeroization of the Mean Bias on the Performance of IOL Power Formulas

Damien Gatinel, Guillaume Debellemanière, Alain Saad, Luis F Brenner, Mathieu Gauvin, Avi Wallerstein, Jacques Malet

Translational Vision Science & Technology · 2024

DOI: 10.1167/tvst.13.10.22 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Translational Vision Science & Technology, 2024. Verfasst von Damien Gatinel und 6 weiteren Autor:innen. Beteiligte Einrichtung: Anterior Segment and Refractive Surgery Department, Rothschild Foundation Hospital, Paris, France. Hauptthemen laut MEDLINE: Lenses, Intraocular, Refraction, Ocular, Optics and Photonics. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

In cataract surgery, accurate intraocular lens (IOL) power calculations are crucial for optimal postoperative refractive outcomes. This study explores the impact of prioritizing the reduction of the standard deviation (SD) of prediction errors before mean prediction error (PE) adjustment on IOL calculation formula precision and accuracy.

Methods

We conducted a retrospective analysis of 4885 eyes from 2611 patients, all implanted with the same IOL model, comparing four traditional IOL power calculation formulas: SRK/T, Holladay 1, Haigis, and Hoffer Q. We introduced new constants aiming to minimize the SD of PE (new_const) against traditionally optimized constants (classic_const), using a heteroscedastic statistical method for comparison. Validation of precision improvements used a secondary dataset of 262 eyes from 132 patients.

Results

We observed significant reductions in mean absolute error (MAE) across training and test sets for Hoffer Q, Holladay, and Haigis formulas, indicating accuracy enhancements. Optimized constants significantly reduced SDs for Haigis from 0.3255 to 0.3153 and for Hoffer Q from 0.3521 to 0.3387. These optimizations also increased the proportion of eyes achieving PE within ±0.25 D. SRK/T showed improved SD from 0.3596 to 0.3585. However, Holladay 1 showed minimal change with no significant improvement. In the test dataset, significant reductions in SD were observed for Haigis and Hoffer Q.

Conclusions

Prioritizing SD minimization before adjusting mean PE significantly improves the precision of selected IOL power formulas, enhancing postoperative refractive outcomes. The effectiveness varies among formulas, underscoring the need for formula-specific adjustments. Translational Relevance: The study presents a novel two-step approach for optimizing IOL power calculations.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Lenses, IntraocularRefraction, OcularOptics and PhotonicsHumansRetrospective StudiesFemaleMaleAgedMiddle AgedLens Implantation, IntraocularVisual AcuityBiometryCataract ExtractionPhacoemulsification

Autor:innen und Institutionen

  • Anterior Segment and Refractive Surgery Department, Rothschild Foundation Hospital, Paris, France.
  • Memira Eye Center, Oslo, Norway.
  • Department of Ophthalmology and Visual Sciences, McGill University, Montreal, Quebec, Canada.
  • LASIK MD, Montreal, Quebec, Canada.

Zitieren

APA

Gatinel, D., Debellemanière, G., Saad, A., Brenner, L. F., Gauvin, M., Wallerstein, A., & Malet, J. (2024). Impact of the Minimization of Standard Deviation Before Zeroization of the Mean Bias on the Performance of IOL Power Formulas. Translational Vision Science & Technology. https://doi.org/10.1167/tvst.13.10.22

Vancouver

Gatinel D, Debellemanière G, Saad A, Brenner LF, Gauvin M, Wallerstein A, et al. Impact of the Minimization of Standard Deviation Before Zeroization of the Mean Bias on the Performance of IOL Power Formulas. Translational Vision Science & Technology. 2024. doi:10.1167/tvst.13.10.22

BibTeX
@article{Gatinel2024,
  author = {Gatinel, Damien and Debellemanière, Guillaume and Saad, Alain and Brenner, Luis F and Gauvin, Mathieu and Wallerstein, Avi and Malet, Jacques},
  title = {Impact of the Minimization of Standard Deviation Before Zeroization of the Mean Bias on the Performance of IOL Power Formulas},
  journal = {Translational Vision Science & Technology},
  year = {2024},
  doi = {10.1167/tvst.13.10.22}
}
Kostenlos starten →