Establishing Fully-Automated Fundus-Controlled Dark Adaptometry: A Validation and Retest-Reliability Study
Translational Vision Science & Technology · 2023
DOI: 10.1167/tvst.12.12.18 ↗Kurz gefasst
Fachartikel in der Zeitschrift Translational Vision Science & Technology, 2023. Verfasst von Jeannine M Oertli und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ophthalmology, University Hospital Basel, Basel, Switzerland. Hauptthemen laut MEDLINE: Retinal Rod Photoreceptor Cells, Retinal Cone Photoreceptor Cells. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 47 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Wellcome Trust.
Abstract (Originalsprache)
Purpose
The purpose of this study was to establish and validate a novel fundus-controlled dark-adaptometry method.
Methods
We developed a custom dark-adaptometry software for the S-MAIA device using the open-perimetry-interface. In the validation-substudy, participants underwent dark-adaptometry testing with a comparator device (MonCvONE, 59% rhodopsin bleach, cyan and red stimuli centered at 2 degrees, 4 degrees, and 6 degrees eccentricity). Following a brief break (approximately 5 minutes), the participants were bleached again and underwent dark-adaptometry testing with the S-MAIA device (same loci). In the retest reliability-substudy, participants were tested twice with the S-MAIA device (same loci as above). Nonlinear curve fitting was applied to extract dark-adaptation curve parameters. Validity and repeatability were summarized in terms of the mean bias and 95% limits of agreement (LoAs).
Results
In the validation-substudy (N = 20 participants, median age interquartile range [IQR] 31.5 years [IQR = 25.8, 62.0]), measures of rod-mediated dark-adaptation showed little to no between method differences for the cone-rod-break-time (bias 95% confidence interval [95% CI] of +0.1 minutes [95% CI = -0.6 to 0.8]), rod-intercept-time (-0.23 minutes [95% CI = -1.38 to 0.93]), and S2 slope (-0.01 LogUnits/minutes [95% CI = -0.02 to -0.01]). In the retest reliability-substudy (N = 10 participants, 32.0 years [95% CI = 27.0, 57.5]), the corresponding LoAs were (cone-rod-break-time) -3.94 to 2.78 minutes, (rod-intercept-time) -4.55 to 3.11 minutes, and (S2 slope [rate-limited component of rod recovery]) -0.03 to 0.03 LogUnits/minutes. The LoAs for the steady-state cone and rod thresholds were -0.28 to 0.33 LogUnits and -0.34 to 0.28 LogUnits.
Conclusions
The devised fundus-controlled dark-adaptometry method yields valid and reliable results. Translational Relevance: Fundus-controlled dark-adaptometry solves the critical need for localized testing of the visual cycle and retinoid transfer in eyes with unstable fixation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Ophthalmology, University Hospital Basel, Basel, Switzerland.
- https://orcid.org/0009-0007-9310-9134.
- https://orcid.org/0000-0002-3642-3502.
- Institute of Molecular and Clinical Ophthalmology Basel, Basel, Switzerland.
- https://orcid.org/0000-0002-7312-0316.
- Ophthalmic Genetics and Visual Function Branch, National Eye Institute, National Institutes of Health, Bethesda, MD, USA.
Förderung
- Wellcome Trust
Zitieren
APA
Oertli, J. M., Pfau, K., Scholl, H. P. N., Jeffrey, B. G., & Pfau, M. (2023). Establishing Fully-Automated Fundus-Controlled Dark Adaptometry: A Validation and Retest-Reliability Study. Translational Vision Science & Technology. https://doi.org/10.1167/tvst.12.12.18
Vancouver
Oertli JM, Pfau K, Scholl HPN, Jeffrey BG, Pfau M. Establishing Fully-Automated Fundus-Controlled Dark Adaptometry: A Validation and Retest-Reliability Study. Translational Vision Science & Technology. 2023. doi:10.1167/tvst.12.12.18
BibTeX
@article{Oertli2023,
author = {Oertli, Jeannine M and Pfau, Kristina and Scholl, Hendrik P N and Jeffrey, Brett G and Pfau, Maximilian},
title = {Establishing Fully-Automated Fundus-Controlled Dark Adaptometry: A Validation and Retest-Reliability Study},
journal = {Translational Vision Science & Technology},
year = {2023},
doi = {10.1167/tvst.12.12.18}
}
EUVIMED