Safe Radiotherapy Delivery in a Pacemaker-dependent Breast Cancer Patient: Insights from an Initial Institutional Experience

Rajiv Singh, Subrata Roy

Journal of Cancer Treatment and Research · 27.05.2026 · 14 (2) · 66-74

DOI: 10.11648/j.jctr.20261402.12 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Cancer Treatment and Research, online erschienen am 27.05.2026. Verfasst von Rajiv Singh und 1 weiteren Autor:innen. Die Arbeit nennt 15 Literaturangaben.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Radiation therapy (RT) is a standard component of breast cancer management; however, treatment planning becomes more complex in patients with cardiac implantable electronic devices (CIEDs), including pacemakers (PMs) and implantable cardioverter-defibrillators (ICDs). Exposure of these devices to therapeutic radiation may lead to malfunction, particularly in pacemaker-dependent individuals, where device failure can have serious clinical consequences. Careful treatment planning is therefore essential when irradiating breast cancer patients with implanted cardiac devices. This report presents our institutional experience in the safe administration of adjuvant RT in a pacemaker-dependent patient undergoing treatment for breast cancer. Materials and Methods: A retrospective review was conducted of five breast cancer patients with dual-chamber pacemakers, all of whom were pacemaker-dependent, treated with adjuvant external beam radiation therapy between January 2023 and December 2024, Simulation imaging was performed using both free-breathing and deep inspiration breath-hold (DIBH) computed tomography (CT) protocols. Target volumes and organs at risk were delineated according to ESTRO/RTOG contouring guidelines, with the pacemaker generator and leads contoured separately as organs at risk. Treatment planning was performed using 6 MV photon beams with field-in-field and hybrid intensity-modulated radiation therapy (IMRT) techniques. Volumetric modulated arc therapy (VMAT) plans were additionally generated for dosimetric comparison. Pacemaker dose constraints were established in accordance with the recommendations of the American Association of Physicists in Medicine Task Group 203 (AAPM TG-203). Prior to treatment initiation, multidisciplinary evaluation was undertaken, including cardiology consultation, device interrogation before and after treatment, and preparation of emergency pacing support protocols.

Results

Adjuvant RT was completed successfully without interruption. Target coverage was clinically acceptable, with 96.8% of the planning target volume (PTV) receiving at least 95% of the prescribed dose (PTV V95% = 96.8%). Use of DIBH contributed to cardiac sparing, resulting in a mean heart dose of 1.9 Gy and a mean dose to the left anterior descending artery (LAD) of 4.1 Gy. The maximum recorded dose to the pacemaker was 0.86 Gy, remaining well below established safety thresholds. No pacemaker malfunction, sensing abnormalities, or cardiac complications were observed during treatment. Acute toxicity was limited to grade 1 radiation dermatitis.

Conclusion

Adjuvant breast radiotherapy in pacemaker-dependent patients can be delivered safely with appropriate precautions. The use of low-energy photon beams, DIBH techniques, meticulous dosimetric planning, and multidisciplinary coordination enables effective target coverage while maintaining pacemaker dose within acceptable limits, without compromising oncologic outcomes or device functionality.

Zitieren

APA

Singh, R., & Roy, S. (2026). Safe Radiotherapy Delivery in a Pacemaker-dependent Breast Cancer Patient: Insights from an Initial Institutional Experience. Journal of Cancer Treatment and Research, 14(2), 66-74. https://doi.org/10.11648/j.jctr.20261402.12

Vancouver

Singh R, Roy S. Safe Radiotherapy Delivery in a Pacemaker-dependent Breast Cancer Patient: Insights from an Initial Institutional Experience. Journal of Cancer Treatment and Research. 2026;14(2):66-74. doi:10.11648/j.jctr.20261402.12

BibTeX
@article{Singh2026,
  author = {Singh, Rajiv and Roy, Subrata},
  title = {Safe Radiotherapy Delivery in a Pacemaker-dependent Breast Cancer Patient: Insights from an Initial Institutional Experience},
  journal = {Journal of Cancer Treatment and Research},
  year = {2026},
  volume = {14},
  number = {2},
  pages = {66--74},
  doi = {10.11648/j.jctr.20261402.12}
}
Kostenlos starten →