A challenging laboratory case due to inappropriate sampling through an antiseptic barrier cap

Giovanni Introcaso, Giorgia Piras, Giorgia Mandelli, Sabrina Frisoli, Fabio Bottari

Biochemia medica · 15.06.2026 · 36 (2) · 369-374

DOI: 10.11613/bm.2026.021001 ↗

Kurz gefasst

Fallbericht in der Zeitschrift Biochemia medica, online erschienen am 15.06.2026. Verfasst von Giovanni Introcaso und 4 weiteren Autor:innen. Beteiligte Einrichtung: Clinical Laboratory Unit, Centro Cardiologico Monzino IRCCS, Milan, Italy. Hauptthemen laut MEDLINE: Anti-Infective Agents, Local, Blood Specimen Collection. Die Arbeit nennt 12 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Antiseptic barrier caps are currently used as novel tools in critically ill patients to achieve sterility of the blood sampling site and to prevent central line-associated bloodstream infections in the intensive care unit. A clinical and laboratory case of analytical interferences on erythrocytes, platelets and leukocytes due to the isopropyl alcohol contained in the antiseptic barrier cap is described. An aged man with history of heart valve disease was monitored for hematological parameters after surgery of mitral and tricuspid valve replacement. Leukocyte scattergrams changes on the Sysmex XN 2000 analyzer with the suspicion of platelet clumps were observed. After investigation on sampling procedure, antiseptic barrier cap of an arterial access (BD PureHub disinfecting cap) was used. Another sample has been requested, and the analysis of complete blood count and peripheral blood smear confirmed the presence of clumped microerythrocytes and erythrocytes with damaged membranes. In addition, differential leukocyte count revealed an overestimation of eosinophils by 10%. A new sample taken from venous access provided accurate results without alterations. The preanalytical error generated during blood sampling highlights the importance of correct use of biomedical devices. Furthermore, hematology analyzer technology represents a promising tool for assessing sample quality.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Anti-Infective Agents, LocalBlood Specimen CollectionHumansMale2-PropanolSpecimen Handling

Stichwörter der Autor:innen

alcohol contaminationanalytical interferenceshematological analyzerinappropriate samplingpreanalytical phase

Substanzen

Anti-Infective Agents, Local2-Propanol

Autor:innen und Institutionen

  • Clinical Laboratory Unit, Centro Cardiologico Monzino IRCCS, Milan, Italy.
  • Intensive Care Unit, Centro Cardiologico Monzino IRCCS, Milan, Italy.

Zitieren

APA

Introcaso, G., Piras, G., Mandelli, G., Frisoli, S., & Bottari, F. (2026). A challenging laboratory case due to inappropriate sampling through an antiseptic barrier cap. Biochemia medica, 36(2), 369-374. https://doi.org/10.11613/bm.2026.021001

Vancouver

Introcaso G, Piras G, Mandelli G, Frisoli S, Bottari F. A challenging laboratory case due to inappropriate sampling through an antiseptic barrier cap. Biochemia medica. 2026;36(2):369-374. doi:10.11613/bm.2026.021001

BibTeX
@article{Introcaso2026,
  author = {Introcaso, Giovanni and Piras, Giorgia and Mandelli, Giorgia and Frisoli, Sabrina and Bottari, Fabio},
  title = {A challenging laboratory case due to inappropriate sampling through an antiseptic barrier cap},
  journal = {Biochemia medica},
  year = {2026},
  volume = {36},
  number = {2},
  pages = {369--374},
  doi = {10.11613/bm.2026.021001}
}
Kostenlos starten →