Practical application of the patient data-based quality control method: the potassium example
Biochemia medica · 15.02.2024 · 34 (1) · 137-141
DOI: 10.11613/bm.2024.010901 ↗Kurz gefasst
Fachartikel in der Zeitschrift Biochemia medica, online erschienen am 15.02.2024. Verfasst von Yan Zhang und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Laboratory, Traditional Chinese Medicine Hospital of Pidu District, Chengdu, China. Hauptthemen laut MEDLINE: Laboratories. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 10 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Introduction
Internal quality control (IQC) is a core pillar of laboratory quality control strategies. Internal quality control commercial materials lack the same characteristics as patient samples and IQC contributes to the costs of laboratory testing. Patient data-based quality control (PDB-QC) may be a valuable supplement to IQC; the smaller the biological variation, the stronger the ability to detect errors. Using the potassium concentration in serum as an example study compared error detection effectiveness between PDB-QC and IQC. Materials and methods: Serum potassium concentrations were measured by using an indirect ion-selective electrode method. For the training database, 23,772 patient-generated data and 366 IQC data from April 2022 to September 2022 were used; 15,351 patient-generated data and 246 IQC data from October 2022 to January 2023 were used as the testing database. For both PDB-QC and IQC, average values and standard deviations were calculated, and z-score charts were plotted for comparison purposes.
Results
Five systematic and three random errors were detected using IQC. Nine systematic errors but no random errors were detected in PDB-QC. The PDB-QC showed systematic error warnings earlier than the IQC.
Conclusions
The daily average value of patient-generated data was superior to IQC in terms of the efficiency and timeliness of detecting systematic errors but inferior to IQC in detecting random errors.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Clinical Laboratory, Traditional Chinese Medicine Hospital of Pidu District, Chengdu, China.
- Department of Clinical Laboratory, The Third Affiliated Hospital of Chengdu University of Chinese Medicine, Chengdu, China.
Zitieren
APA
Zhang, Y., Wang, H. L., Xie, Y. H., He, D. H., Zhou, C. Q., & Kong, L. R. (2024). Practical application of the patient data-based quality control method: the potassium example. Biochemia medica, 34(1), 137-141. https://doi.org/10.11613/bm.2024.010901
Vancouver
Zhang Y, Wang HL, Xie YH, He DH, Zhou CQ, Kong LR. Practical application of the patient data-based quality control method: the potassium example. Biochemia medica. 2024;34(1):137-141. doi:10.11613/bm.2024.010901
BibTeX
@article{Zhang2024,