Analytical bias of automated immunoassays for six serum steroid hormones assessed by LC-MS/MS

Željko Debeljak, Ivana Marković, Jasna Pavela, Iva Lukić, Dario Mandić, Sanja Mandić, Vesna Horvat, Vatroslav Šerić

Biochemia medica · 15.10.2020 · 30 (3) · 422-431

DOI: 10.11613/bm.2020.030701 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biochemia medica, online erschienen am 12.10.2020. Verfasst von Željko Debeljak und 7 weiteren Autor:innen. Beteiligte Einrichtung: Clinical Institute of Laboratory Diagnostics, Osijek University Hospital, Osijek, Croatia. Hauptthemen laut MEDLINE: Chromatography, High Pressure Liquid, Immunoassay, Tandem Mass Spectrometry. Die Arbeit wurde bisher 25-mal zitiert (Crossref) und nennt 23 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

There is a growing amount of evidence showing the significant analytical bias of steroid hormone immunoassays, but large number of available immunoassays makes conduction of a single comprehensive study of this issue hardly feasible. Aim of this study was to assess the analytical bias of six heterogeneous immunoassays for serum aldosterone, cortisol, dehydroepiandrosterone sulphate (DHEAS), testosterone, 17-hydroxyprogesterone (OHP) and progesterone using the liquid chromatography coupled to the tandem mass spectrometry (LC-MS/MS). Materials and methods: This method comparison study included 49 serum samples. Testosterone, DHEAS, progesterone and cortisol immunoassays were performed on the Abbott Architect i2000SR or Alinity i analysers (Abbott Diagnostics, Chicago, USA). DiaSorin’s Liaison (DiaSorin, Saluggia, Italy) and DIAsource’s ETI-Max 3000 analysers (DIAsource ImmunoAssays, Louvain-La-Neuve, Belgium) were chosen for aldosterone and OHP immunoassay testing, respectively. All immunoassays were evaluated against the LC-MS/MS assay relying on the commercial kit (Chromsystems, Gräfelfing, Germany) and LCMS-8050 analyser (Shimadzu, Kyoto, Japan). Analytical biases were calculated and method comparison was conducted using weighted Deming regression analysis.

Results

Depending on the analyte and specific immunoassay, mean relative biases ranged from -31 to + 137%. Except for the cortisol, immunoassays were positively biased. For none of the selected steroids slope and intercept 95% confidence intervals simultaneously contained 0 and 1, respectively.

Conclusions

Evaluated immunoassays failed to satisfy requirements for methods’ comparability and produced significant analytical biases in respect to the LC-MS/MS assay, especially at low concentrations.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Chromatography, High Pressure LiquidImmunoassayTandem Mass Spectrometry17-alpha-HydroxyprogesteroneAdultAldosteroneAutomationBiasDehydroepiandrosterone SulfateFemaleHumansHydrocortisoneMaleMiddle AgedProgesteroneReagent Kits, DiagnosticSteroidsTestosteroneYoung Adult

Stichwörter der Autor:innen

LC-MS/MSbiasimmunoassaymethod comparisonsteroid hormone

Substanzen

Reagent Kits, DiagnosticSteroidsTestosteroneAldosteroneProgesteroneDehydroepiandrosterone Sulfate17-alpha-HydroxyprogesteroneHydrocortisone

Autor:innen und Institutionen

  • Clinical Institute of Laboratory Diagnostics, Osijek University Hospital, Osijek, Croatia.
  • Faculty of Medicine, JJ Strossmayer University of Osijek, Osijek, Croatia.

Zitieren

APA

Debeljak, Ž., Marković, I., Pavela, J., Lukić, I., Mandić, D., Mandić, S., Horvat, V., & Šerić, V. (2020). Analytical bias of automated immunoassays for six serum steroid hormones assessed by LC-MS/MS. Biochemia medica, 30(3), 422-431. https://doi.org/10.11613/bm.2020.030701

Vancouver

Debeljak Ž, Marković I, Pavela J, Lukić I, Mandić D, Mandić S, et al. Analytical bias of automated immunoassays for six serum steroid hormones assessed by LC-MS/MS. Biochemia medica. 2020;30(3):422-431. doi:10.11613/bm.2020.030701

BibTeX
@article{Debeljak2020,
  author = {Debeljak, Željko and Marković, Ivana and Pavela, Jasna and Lukić, Iva and Mandić, Dario and Mandić, Sanja and Horvat, Vesna and Šerić, Vatroslav},
  title = {Analytical bias of automated immunoassays for six serum steroid hormones assessed by LC-MS/MS},
  journal = {Biochemia medica},
  year = {2020},
  volume = {30},
  number = {3},
  pages = {422--431},
  doi = {10.11613/bm.2020.030701}
}
Kostenlos starten →