Introduction/Purpose: Emerging evidence suggests that distal penetration of the embolic may provide faster resolution of the subdural hematoma. The distal penetration of a liquid embolic device under development is compared against existing embolic devices in a pre-clinical model. Materials/Methods: The test article (IMPASS Embolic Device) was compared to TRUFILL n-BCA diluted 1:5 with Lipiodol (control 1) and Onyx-18 (control 2) using a swine kidney model including two Yorkshire cross domestic swine. Test and control articles were delivered via microcatheter into distinct regions of the renal arterial vasculature by a neurovascular interventionalist. Following surgery, animals were euthanized, and the kidneys were fixed in 10% neutral buffered formalin and evaluated by a board-certified veterinary pathologist. The distal penetration of test articles was characterized by the pathologist using a rubric (Figure 1) and three different methods 1) completeness of vessel filling, 2) frequency of completeness of filling, and 3) minimum vessel size where embolic was observed. For method one (1) and two (2) the kidney sections were divided into three anatomically based zones starting from the hilus (Zone 1) and extending to the distal cortex (Zone 3). Given the general inversely proportional relationship between arterial size and distance from the hilum, these zones were determined to reflect overall distal penetration of the products considered.
Results
The test article exhibited an average arterial occlusion score of 3.6, compared to 2.6 for control 1 and 2.6 for control 2, where 4 equals complete occlusion of the lumen. The test article showed complete occlusion (Score 4) in 76%, 72%, 67% of treated arteries in Zone 1, 2, and 3 respectively, in comparison control 1 exhibited 14%, 0%, 0% respectively and control 2 exhibited 12%, 0%, and 29% respectively (Figure 1). Finally, the smallest diameter vessels where embolic materials were observed was found to be 9.9 µm, 47.7 µm, 34.6 µm for the test article, control 1, and control 2 respectively.
Conclusion
The test article consistently scored higher with respect to complete vessel filling, frequency of complete vessel filling, and smallest vessel diameter reached. The test article may be a suitable embolic for MMA embolization and other applications where deep distal reach is desired.
Zitieren
APA
Grandhi, R., Smith, D., Bjorklund, R., Bainsmith, M., & Hodgson, K. (2026). Abstract 1118: Distal Penetration Assessment of Liquid Embolic Materials in a Pre-Clinical Model. Stroke: Vascular and Interventional Neurology, 6(Suppl_1). https://doi.org/10.1161/svi270000_1118
Vancouver
Grandhi R, Smith D, Bjorklund R, Bainsmith M, Hodgson K. Abstract 1118: Distal Penetration Assessment of Liquid Embolic Materials in a Pre-Clinical Model. Stroke: Vascular and Interventional Neurology. 2026;6(Suppl_1). doi:10.1161/svi270000_1118
BibTeX
@article{Grandhi2026,
author = {Grandhi, R. and Smith, D. and Bjorklund, R. and Bainsmith, M. and Hodgson, K.},
title = {Abstract 1118: Distal Penetration Assessment of Liquid Embolic Materials in a Pre-Clinical Model},
journal = {Stroke: Vascular and Interventional Neurology},
year = {2026},
volume = {6},
number = {Suppl_1},
doi = {10.1161/svi270000_1118}
}
EUVIMED