Fachartikel in der Zeitschrift Journal of the American Heart Association, 2024. Verfasst von Hyun Jung Koo und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology and Research Institute of Radiology, Cardiac Imaging Center University of Ulsan College of Medicine, Asan Medical Center Seou…. Hauptthemen laut MEDLINE: Transcatheter Aortic Valve Replacement, Hemodynamics, Heart Valve Prosthesis, Coronary Circulation. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 23 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background This study aimed to investigate the hemodynamic and anatomic factors associated with sinus thrombosis following transcatheter aortic valve replacement (TAVR), integrating in vivo patient data analysis and in vitro experiments. Methods and Results Postprocedural, 4‐dimensional, multiphase computed tomography data from 211 patients enrolled in the ADAPT‐TAVR (Anticoagulation Versus Dual Antiplatelet Therapy for Prevention of Leaflet Thrombosis and Cerebral Embolization After Transcatheter Aortic Valve Replacement) study were analyzed. The prevalence of native sinus thrombosis was examined in relation to valve type, implant depth, and anatomic features. In vitro experiments used particle image velocimetry to observe changes in sinus flow based on the transcatheter heart valves (23‐mm SAPIEN3, Edwards Lifesciences; and 29‐mm CoreValve, Medtronic) height and coronary artery flow. Native sinus thrombosis was more common in self‐expanding valves (39.1% versus 14.9%, P =0.004). In per‐cusp analysis of in vivo patient data, adjusted transcatheter heart valve implant depth (odds ratio, 1.2 [95% CI, 1.1–1.3]; P

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Transcatheter Aortic Valve ReplacementHemodynamicsHeart Valve ProsthesisCoronary CirculationAortic ValveHumansFemaleMaleAged, 80 and overAgedThrombosisProsthesis DesignAortic Valve StenosisSinus of ValsalvaRisk FactorsAnticoagulantsMultidetector Computed TomographyPlatelet Aggregation Inhibitors

Stichwörter der Autor:innen

aortic stenosiscomputed tomographythrombosistranscatheter aortic valve replacement

Substanzen

AnticoagulantsPlatelet Aggregation Inhibitors

Autor:innen und Institutionen

  • Department of Radiology and Research Institute of Radiology, Cardiac Imaging Center University of Ulsan College of Medicine, Asan Medical Center Seoul South Korea.
  • Interdisciplinary Program in Biohealth-Machinery Convergence Engineering Kangwon National University Chuncheon South Korea.
  • Division of Cardiology University of Ulsan College of Medicine, Asan Medical Center Seoul South Korea.
  • Center for Precision Medicine Platform Based on Smart Hemo-Dynamic Index Seoul South Korea.

ORCID-Profile: Hyun Jung Koo, Jihun Kang, Do-Yoon Kang, Jung-Min Ahn, Duk-Woo Park, Seung-Jung Park, Joon-Won Kang, Hojin Ha, Dong Hyun Yang

Zitieren

APA

Koo, H. J., Kang, J., Kang, D. Y., Ahn, J. M., Park, D. W., Park, S. J., Kang, J. W., Ha, H., & Yang, D. H. (2024). Native Sinus Hemodynamics and Thrombosis in Transcatheter Heart Valve: Effect of Implant Depth and Coronary Flow. Journal of the American Heart Association. https://doi.org/10.1161/jaha.124.037105

Vancouver

Koo HJ, Kang J, Kang DY, Ahn JM, Park DW, Park SJ, et al. Native Sinus Hemodynamics and Thrombosis in Transcatheter Heart Valve: Effect of Implant Depth and Coronary Flow. Journal of the American Heart Association. 2024. doi:10.1161/jaha.124.037105

BibTeX
@article{Koo2024,
  author = {Koo, Hyun Jung and Kang, Jihun and Kang, Do-Yoon and Ahn, Jung-Min and Park, Duk-Woo and Park, Seung-Jung and Kang, Joon-Won and Ha, Hojin and Yang, Dong Hyun},
  title = {Native Sinus Hemodynamics and Thrombosis in Transcatheter Heart Valve: Effect of Implant Depth and Coronary Flow},
  journal = {Journal of the American Heart Association},
  year = {2024},
  doi = {10.1161/jaha.124.037105}
}
Kostenlos starten →