Global Results of Implantable Loop Recorder for Detection of Atrial Fibrillation After Stroke: Reveal LINQ Registry

Kazunori Toyoda iD, Kengo Kusano iD, Yasuyuki Iguchi iD, Takanori Ikeda, Itsuro Morishima iD, Hirofumi Tomita iD, Taku Asano iD, Teiichi Yamane iD, Ichiro Nakahara iD, Eiichi Watanabe, Junjiroh Koyama iD, Ritsushi Kato iD und 15 weitere

Journal of the American Heart Association · 2024

DOI: 10.1161/jaha.124.035956 ↗

Kurz gefasst

Beobachtungsstudie in der Zeitschrift Journal of the American Heart Association, 2024. Verfasst von Kazunori Toyoda und 26 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cerebrovascular Medicine National Cerebral and Cardiovascular Center Suita Japan. Hauptthemen laut MEDLINE: Atrial Fibrillation, Registries, Electrocardiography, Ambulatory, Ischemic Stroke. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 48 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background We aimed to quantify the incidence of atrial fibrillation (AF) in patients with cryptogenic stroke globally, as well as separately in patients in and outside of Japan, using an implantable loop recorder from a prospective, observational, Reveal LINQ Registry. Methods and Results Patients developing cryptogenic stroke and monitored by implantable loop recorder for searching AF were studied. The primary end point was incidence of AF within 36 months after insertion. Secondary end points were recurrent ischemic stroke/transient ischemic attack and AF‐related treatment strategies. A total of 271 patients (61.6±14.3 years, 170 men, 60 from Japan) were enrolled from 12 countries. AF was detected in 28.2% at 36 months. The median time from enrollment to AF detection was 7.9 months. During the first 12 months, the AF detection rate slope was relatively steeper in the Japanese subgroup versus non‐Japanese patients. However, by 3 years, the cumulative incidence of AF detection did not differ between groups. Age was the only variable associated with AF detection (hazard ratio, 1.05 [95% CI, 1.02–1.07] per year), trending higher in older age groups. Of the 271 patients, 11 (4.1%) developed recurrent ischemic stroke/transient ischemic attack; AF was detected by implantable loop recorder in only 1 of these patients. Patients with detected AF were more commonly taking oral anticoagulation than those without AF at the last follow‐up (64.7% versus 25.3%, P

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Atrial FibrillationRegistriesElectrocardiography, AmbulatoryIschemic StrokeHumansMaleFemaleMiddle AgedAgedIncidenceProspective StudiesRecurrenceTime FactorsRisk FactorsIschemic Attack, TransientStrokeJapan

Stichwörter der Autor:innen

anticoagulationatrial fibrillationcardioembolismcerebral infarctioncryptogenic ischemic strokeelectrocardiographic monitoring

Autor:innen und Institutionen

  • Department of Cerebrovascular Medicine National Cerebral and Cardiovascular Center Suita Japan.
  • Department of Cardiovascular Medicine National Cerebral and Cardiovascular Center Suita Japan.
  • Department of Neurology The Jikei University School of Medicine Tokyo Japan.
  • Department of Cardiovascular Medicine Toho University Faculty of Medicine Tokyo Japan.
  • Department of Cardiology Ogaki Municipal Hospital Ogaki Japan.
  • Department of Cardiology and Nephrology Hirosaki University Graduate School of Medicine Hirosaki Japan.

ORCID-Profile: Kazunori Toyoda, Kengo Kusano, Yasuyuki Iguchi, Itsuro Morishima, Hirofumi Tomita, Taku Asano, Teiichi Yamane, Ichiro Nakahara, Junjiroh Koyama, Ritsushi Kato, Hiroshi Morita, Teruyuki Hirano, Kyoko Soejima, Shingen Owada, Haruhiko Abe, Masahiro Yasaka, Toshihiro Nakamura, Scott Kasner, Andrea Natale, Alpesh N Amin, Erika Pouliot, Noreli Franco, Kazuhiro Hidaka, Ken Okumura

Zitieren

APA

Toyoda, K., Kusano, K., Iguchi, Y., Ikeda, T., Morishima, I., Tomita, H., Asano, T., Yamane, T., Nakahara, I., Watanabe, E., Koyama, J., Kato, R., Morita, H., Hirano, T., Soejima, K., Owada, S., Abe, H., Yasaka, M., Nakamura, T., …, & Okumura, K. (2024). Global Results of Implantable Loop Recorder for Detection of Atrial Fibrillation After Stroke: Reveal LINQ Registry. Journal of the American Heart Association. https://doi.org/10.1161/jaha.124.035956

Vancouver

Toyoda K, Kusano K, Iguchi Y, Ikeda T, Morishima I, Tomita H, et al. Global Results of Implantable Loop Recorder for Detection of Atrial Fibrillation After Stroke: Reveal LINQ Registry. Journal of the American Heart Association. 2024. doi:10.1161/jaha.124.035956

BibTeX
@article{Toyoda2024,
  author = {Toyoda, Kazunori and Kusano, Kengo and Iguchi, Yasuyuki and Ikeda, Takanori and Morishima, Itsuro and Tomita, Hirofumi and Asano, Taku and Yamane, Teiichi and Nakahara, Ichiro and Watanabe, Eiichi and Koyama, Junjiroh and Kato, Ritsushi and Morita, Hiroshi and Hirano, Teruyuki and Soejima, Kyoko and Owada, Shingen and Abe, Haruhiko and Yasaka, Masahiro and Nakamura, Toshihiro and Kasner, Scott and Natale, Andrea and Beinart, Sean and Amin, Alpesh N and Pouliot, Erika and Franco, Noreli and Hidaka, Kazuhiro and Okumura, Ken},
  title = {Global Results of Implantable Loop Recorder for Detection of Atrial Fibrillation After Stroke: Reveal LINQ Registry},
  journal = {Journal of the American Heart Association},
  year = {2024},
  doi = {10.1161/jaha.124.035956}
}
Kostenlos starten →