Validity and Accuracy of the Derived Left Ventricular End-Diastolic Pressure in Impella 5.5

Reza Poyanmehr iD, Jasmin S Hanke, Dietmar Boethig, Ali Saad Merzah iD, Jan Karsten iD, Paul Frank, Martin Hinteregger iD, Alina Zubarevich, Günes Dogan iD, Jan D Schmitto, Andreas Schäfer iD, L Christian Napp iD und 5 weitere

Circulation: Heart Failure · 2025

DOI: 10.1161/circheartfailure.124.012154 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift Circulation: Heart Failure, 2025. Verfasst von Reza Poyanmehr und 16 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiothoracic, Transplantation and Vascular Surgery (R.P., J.S.H., D.B., A.S.M., M.H., A.Z., G.D., J.D.S., A.F.P., A.W., A.R., B.S.), …. Hauptthemen laut MEDLINE: Heart-Assist Devices, Ventricular Function, Left, Heart Failure, Ventricular Pressure. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 42 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Consensus regarding on-support evaluation and weaning concepts from Impella 5.5 support is scarce. The derived left ventricular end-diastolic pressure (dLVEDP), estimated by device algorithms, is a rarely reported tool for monitoring the weaning process. Its validation and clinical accuracy have not been studied in patients. We assess dLVEDP’s accuracy in predicting pulmonary capillary wedge pressure (PCWP) and propose a corrective equation. METHODS: We included 29 consecutive patients treated with Impella 5.5: 12 in a generation cohort and 17 in a validation cohort. dLVEDP and PCWP were measured 5-fold every 8 hours during support, totaling 698 series with 3490 measurements. Variables such as Impella 5.5 performance level, heart rhythm, pacemaker settings, sex, mechanical ventilation, and body mass index were recorded. Linear regression was used to correct dLVEDP-PCWP discrepancies. Analysis included Bland-Altman plots, linear regression, histograms, and violin plots. RESULTS: The raw dLVEDP and PCWP data did not coincide satisfactorily. The Impella 5.5 dLVEDP overestimation was 3.5±1.5 mm Hg (mean±SD), increasing with higher pressures and unaffected by cardiac rhythm, mechanical ventilation, and performance levels. Statistical correction using the formula modified dLVEDP=−0.457+(1−sex[1=male, 0=female])×0.719−0.0496× body mass index+1.015×body surface area+0.811×dLVEDP significantly reduced the overestimation ( P

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Heart-Assist DevicesVentricular Function, LeftHeart FailureVentricular PressureHumansMaleFemaleAgedMiddle AgedReproducibility of ResultsPulmonary Wedge PressureAlgorithmsPredictive Value of TestsAged, 80 and over

Stichwörter der Autor:innen

heart failurehemodynamicspostoperative careshock, cardiogenicweaning

Autor:innen und Institutionen

  • Department of Cardiothoracic, Transplantation and Vascular Surgery (R.P., J.S.H., D.B., A.S.M., M.H., A.Z., G.D., J.D.S., A.F.P., A.W., A.R., B.S.), Hannover Medical School, Germany.
  • Department of Anesthesiology and Intensive Care Medicine (J.K.), Hannover Medical School, Germany.
  • Department of Anesthesiology and Intensive Care Medicine, Helios Kliniken Mittelweser, Nienburg, Germany (P.F.).
  • Department of Cardiology and Angiology (A.S., L.C.N., J.B.), Hannover Medical School, Germany.

ORCID-Profile: Reza Poyanmehr, Ali Saad Merzah, Jan Karsten, Martin Hinteregger, Günes Dogan, Andreas Schäfer, L Christian Napp, Aron Frederik Popov, Johann Bauersachs

Zitieren

APA

Poyanmehr, R., Hanke, J. S., Boethig, D., Merzah, A. S., Karsten, J., Frank, P., Hinteregger, M., Zubarevich, A., Dogan, G., Schmitto, J. D., Schäfer, A., Napp, L. C., Popov, A. F., Weymann, A., Bauersachs, J., Ruhparwar, A., & Schmack, B. (2025). Validity and Accuracy of the Derived Left Ventricular End-Diastolic Pressure in Impella 5.5. Circulation: Heart Failure. https://doi.org/10.1161/circheartfailure.124.012154

Vancouver

Poyanmehr R, Hanke JS, Boethig D, Merzah AS, Karsten J, Frank P, et al. Validity and Accuracy of the Derived Left Ventricular End-Diastolic Pressure in Impella 5.5. Circulation: Heart Failure. 2025. doi:10.1161/circheartfailure.124.012154

BibTeX
@article{Poyanmehr2025,
  author = {Poyanmehr, Reza and Hanke, Jasmin S and Boethig, Dietmar and Merzah, Ali Saad and Karsten, Jan and Frank, Paul and Hinteregger, Martin and Zubarevich, Alina and Dogan, Günes and Schmitto, Jan D and Schäfer, Andreas and Napp, L Christian and Popov, Aron Frederik and Weymann, Alexander and Bauersachs, Johann and Ruhparwar, Arjang and Schmack, Bastian},
  title = {Validity and Accuracy of the Derived Left Ventricular End-Diastolic Pressure in Impella 5.5},
  journal = {Circulation: Heart Failure},
  year = {2025},
  doi = {10.1161/circheartfailure.124.012154}
}
Kostenlos starten →