Background
Hypertension is a leading preventable cause of morbidity and mortality worldwide. While guidelines recommend out-of-office blood pressure (BP) monitoring, conventional oscillometric methods are intermittent and unsuitable for widespread ambulatory use. Optical sensors offer wearable solutions for continuous BP monitoring and may support hypertension management. While most approaches rely on photoplethysmography (PPG), laser Doppler flowmetry (LDF) is underexplored. We applied a novel approach combining pulse wave analysis (PWA) with machine learning (ML) to both signals, extracting over 99 time- and frequency-domain features for BP estimation.
Objective
To evaluate the accuracy, agreement, and trending ability of PWA-ML-based BP estimation using PPG and LDF, compared to oscillometric reference measurements.
Methods
In a prospective observational study, 49 adults (median age 40 [IQR 25-65], range 21-77 years) underwent two measurement sessions up to three weeks apart, comprising rest and low-intensity exercise. PPG, LDF, and oscillometric measurements were obtained simultaneously on the upper arm and processed using similar algorithms with per-subject calibration to estimate systolic BP (SBP) and diastolic BP (DBP). Performance was evaluated using correlation, Bland-Altman, four-quadrant, and polar plot analyses, with subgroup analyses for subjects aged
Zitieren
APA
Florax, A., Fresiello, L., Kilic, Y., Fransen, A., & Donker, D. (2026). Abstract 6975: Laser Doppler Flowmetry Outperforms Photoplethysmography for Machine Learning-Based Cuffless Blood Pressure Estimation - Validation Against Oscillometry. Circulation, 154(Suppl_1). https://doi.org/10.1161/circ.154.suppl_1.6975
Vancouver
Florax A, Fresiello L, Kilic Y, Fransen A, Donker D. Abstract 6975: Laser Doppler Flowmetry Outperforms Photoplethysmography for Machine Learning-Based Cuffless Blood Pressure Estimation - Validation Against Oscillometry. Circulation. 2026;154(Suppl_1). doi:10.1161/circ.154.suppl_1.6975
BibTeX
@article{Florax2026,
author = {Florax, Anna and Fresiello, Libera and Kilic, Yakup and Fransen, Alwin and Donker, Dirk},
title = {Abstract 6975: Laser Doppler Flowmetry Outperforms Photoplethysmography for Machine Learning-Based Cuffless Blood Pressure Estimation - Validation Against Oscillometry},
journal = {Circulation},
year = {2026},
volume = {154},
number = {Suppl_1},
doi = {10.1161/circ.154.suppl_1.6975}
}
EUVIMED