Abstract 4529: Adverse Events Associated with Da Vinci Xi-Assisted Anatomical Lung Resections: A MAUDE Database Analysis of Lobectomy and Segmentectomy
Unerwünschte Ereignisse bei Da Vinci Xi-unterstützten anatomischen Lungenresektionen: Analyse der MAUDE-Datenbank zu Lobektomie und Segmentektomie
Circulation · 02.11.2026 · 154 (Suppl_1)
DOI: 10.1161/circ.154.suppl_1.4529 ↗Auf einen Blick
In der MAUDE-Datenbank wurden bei Da Vinci Xi-unterstützter Lobektomie und Segmentektomie relevante Geräteprobleme und patientenbezogene Ereignisse beschrieben. Am häufigsten war Blutung/Hämorrhagie.
Fragestellung
Welche gerätebezogenen Probleme und patientenbezogenen unerwünschten Ereignisse sind in der FDA-MAUDE-Datenbank zu Da Vinci Xi-unterstützter pulmonaler Lobektomie und Segmentektomie erfasst?
Methode
Retrospektive Analyse der FDA-MAUDE-Datenbank für Berichte zu Da Vinci Xi-unterstützter Lobektomie und Segmentektomie vom 1. April 2015 bis 19. März 2026; deskriptive Auswertung und Co-Occurrence-Analyse.
Ergebnisse
Von 9.544 gescreenten Berichten erfüllten 315 die Kriterien für Lobektomie und 92 für Segmentektomie. Blutung/Hämorrhagie war das häufigste patientenbezogene Ereignis: 32,9 % bei Lobektomie und 20,4 % bei Segmentektomie.
Schlussfolgerung der Autor:innen
Die Autor:innen berichten bedeutsame gerätebezogene und patientenbezogene Komplikationen bei Da Vinci Xi-unterstützter Lungenresektion. Als wiederkehrende Probleme nennen sie Fehlermeldungen und unbeabsichtigte Systembewegungen; Blutung blieb die häufigste patientenbezogene Komplikation.
Kernaussagen
- 315 Lobektomie- und 92 Segmentektomie-Berichte erfüllten die Einschlusskriterien
- Blutung/Hämorrhagie war das häufigste patientenbezogene Ereignis
- Fehlermeldungen und unbeabsichtigte Systembewegungen traten wiederholt auf
Deutsche Zusammenfassung: KI-gestützt aus dem Original-Abstract erstellt; alle Zahlen wurden automatisch mit dem Original abgeglichen. Maßgeblich ist der Original-Abstract.
Kurz gefasst
Fachartikel in der Zeitschrift Circulation, erschienen am 02.11.2026. Verfasst von Bara Hammadeh.
Themen: Patientensicherheit
Abstract (Originalsprache)
Background
Anatomical lung resections (lobectomy and segmentectomy) are increasingly performed using robotic-assisted thoracic surgery (RATS), particularly with the Da Vinci Xi system. However, postmarketing safety data on device-related adverse events remain limited. Research Question: What are the device-related problems and patient-related adverse events associated with Da Vinci Xi-assisted pulmonary lobectomy and segmentectomy reported to the FDA MAUDE database?
Methods
A retrospective analysis of the FDA MAUDE database was conducted for reports involving Da Vinci Xi-assisted lobectomy and segmentectomy from April 1, 2015, to March 19, 2026. Reports were manually screened for procedural relevance. Device-related problems and patient-related adverse events were extracted, categorized, and analyzed descriptively. Co-occurrence patterns were also examined.
Results
Of 9,544 screened Da Vinci Xi reports, 315 met inclusion for lobectomy (961 device-related events; 5,676 patient-related events) and 92 for segmentectomy (100 device-related events; 96 patient-related events). The most common device-related problems in lobectomy were adverse events without identified device or use problem (28.9%), visual prompts failing to clear (19.2%), and output problems (6.5%). In segmentectomy, visual prompts failing to clear (27.1%) and adverse events without identified device or use problem (27.1%) predominated, followed by unintended system motion (10.4%). Among patient-related events, hemorrhage/bleeding was the most frequent complication in both lobectomy (32.9%) and segmentectomy (20.4%). Death was reported in 7 lobectomy cases (4.3%) and 1 segmentectomy case (2.0%). Co-occurrence analysis revealed visual prompt failures frequently accompanied unintended system motion in segmentectomy.
Conclusion
Significant device-related and patient-related complications are associated with Da Vinci Xi-assisted pulmonary lobectomy and segmentectomy. Visual prompt failures and unintended system motion represent recurring device-related challenges, while hemorrhage remains the predominant patient-related complication. These findings underscore the need for improved user interface design, enhanced surgeon training, and continued postmarketing surveillance to optimize patient safety in robotic-assisted thoracic surgery.
Zitieren
APA
Hammadeh, B. (2026). Abstract 4529: Adverse Events Associated with Da Vinci Xi-Assisted Anatomical Lung Resections: A MAUDE Database Analysis of Lobectomy and Segmentectomy. Circulation, 154(Suppl_1). https://doi.org/10.1161/circ.154.suppl_1.4529
Vancouver
Hammadeh B. Abstract 4529: Adverse Events Associated with Da Vinci Xi-Assisted Anatomical Lung Resections: A MAUDE Database Analysis of Lobectomy and Segmentectomy. Circulation. 2026;154(Suppl_1). doi:10.1161/circ.154.suppl_1.4529
BibTeX
@article{Hammadeh2026,
author = {Hammadeh, Bara},
title = {Abstract 4529: Adverse Events Associated with Da Vinci Xi-Assisted Anatomical Lung Resections: A MAUDE Database Analysis of Lobectomy and Segmentectomy},
journal = {Circulation},
year = {2026},
volume = {154},
number = {Suppl_1},
doi = {10.1161/circ.154.suppl_1.4529}
}