Abstract 4529: Adverse Events Associated with Da Vinci Xi-Assisted Anatomical Lung Resections: A MAUDE Database Analysis of Lobectomy and Segmentectomy

Unerwünschte Ereignisse bei Da Vinci Xi-unterstützten anatomischen Lungenresektionen: Analyse der MAUDE-Datenbank zu Lobektomie und Segmentektomie

Bara Hammadeh

Circulation · 02.11.2026 · 154 (Suppl_1)

DOI: 10.1161/circ.154.suppl_1.4529 ↗

Auf einen Blick

In der MAUDE-Datenbank wurden bei Da Vinci Xi-unterstützter Lobektomie und Segmentektomie relevante Geräteprobleme und patientenbezogene Ereignisse beschrieben. Am häufigsten war Blutung/Hämorrhagie.

Fragestellung

Welche gerätebezogenen Probleme und patientenbezogenen unerwünschten Ereignisse sind in der FDA-MAUDE-Datenbank zu Da Vinci Xi-unterstützter pulmonaler Lobektomie und Segmentektomie erfasst?

Methode

Retrospektive Analyse der FDA-MAUDE-Datenbank für Berichte zu Da Vinci Xi-unterstützter Lobektomie und Segmentektomie vom 1. April 2015 bis 19. März 2026; deskriptive Auswertung und Co-Occurrence-Analyse.

Ergebnisse

Von 9.544 gescreenten Berichten erfüllten 315 die Kriterien für Lobektomie und 92 für Segmentektomie. Blutung/Hämorrhagie war das häufigste patientenbezogene Ereignis: 32,9 % bei Lobektomie und 20,4 % bei Segmentektomie.

Schlussfolgerung der Autor:innen

Die Autor:innen berichten bedeutsame gerätebezogene und patientenbezogene Komplikationen bei Da Vinci Xi-unterstützter Lungenresektion. Als wiederkehrende Probleme nennen sie Fehlermeldungen und unbeabsichtigte Systembewegungen; Blutung blieb die häufigste patientenbezogene Komplikation.

Kernaussagen

  • 315 Lobektomie- und 92 Segmentektomie-Berichte erfüllten die Einschlusskriterien
  • Blutung/Hämorrhagie war das häufigste patientenbezogene Ereignis
  • Fehlermeldungen und unbeabsichtigte Systembewegungen traten wiederholt auf

Deutsche Zusammenfassung: KI-gestützt aus dem Original-Abstract erstellt; alle Zahlen wurden automatisch mit dem Original abgeglichen. Maßgeblich ist der Original-Abstract.

Kurz gefasst

Fachartikel in der Zeitschrift Circulation, erschienen am 02.11.2026. Verfasst von Bara Hammadeh.

Auf einen BlickAbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Anatomical lung resections (lobectomy and segmentectomy) are increasingly performed using robotic-assisted thoracic surgery (RATS), particularly with the Da Vinci Xi system. However, postmarketing safety data on device-related adverse events remain limited. Research Question: What are the device-related problems and patient-related adverse events associated with Da Vinci Xi-assisted pulmonary lobectomy and segmentectomy reported to the FDA MAUDE database?

Methods

A retrospective analysis of the FDA MAUDE database was conducted for reports involving Da Vinci Xi-assisted lobectomy and segmentectomy from April 1, 2015, to March 19, 2026. Reports were manually screened for procedural relevance. Device-related problems and patient-related adverse events were extracted, categorized, and analyzed descriptively. Co-occurrence patterns were also examined.

Results

Of 9,544 screened Da Vinci Xi reports, 315 met inclusion for lobectomy (961 device-related events; 5,676 patient-related events) and 92 for segmentectomy (100 device-related events; 96 patient-related events). The most common device-related problems in lobectomy were adverse events without identified device or use problem (28.9%), visual prompts failing to clear (19.2%), and output problems (6.5%). In segmentectomy, visual prompts failing to clear (27.1%) and adverse events without identified device or use problem (27.1%) predominated, followed by unintended system motion (10.4%). Among patient-related events, hemorrhage/bleeding was the most frequent complication in both lobectomy (32.9%) and segmentectomy (20.4%). Death was reported in 7 lobectomy cases (4.3%) and 1 segmentectomy case (2.0%). Co-occurrence analysis revealed visual prompt failures frequently accompanied unintended system motion in segmentectomy.

Conclusion

Significant device-related and patient-related complications are associated with Da Vinci Xi-assisted pulmonary lobectomy and segmentectomy. Visual prompt failures and unintended system motion represent recurring device-related challenges, while hemorrhage remains the predominant patient-related complication. These findings underscore the need for improved user interface design, enhanced surgeon training, and continued postmarketing surveillance to optimize patient safety in robotic-assisted thoracic surgery.

Zitieren

APA

Hammadeh, B. (2026). Abstract 4529: Adverse Events Associated with Da Vinci Xi-Assisted Anatomical Lung Resections: A MAUDE Database Analysis of Lobectomy and Segmentectomy. Circulation, 154(Suppl_1). https://doi.org/10.1161/circ.154.suppl_1.4529

Vancouver

Hammadeh B. Abstract 4529: Adverse Events Associated with Da Vinci Xi-Assisted Anatomical Lung Resections: A MAUDE Database Analysis of Lobectomy and Segmentectomy. Circulation. 2026;154(Suppl_1). doi:10.1161/circ.154.suppl_1.4529

BibTeX
@article{Hammadeh2026,
  author = {Hammadeh, Bara},
  title = {Abstract 4529: Adverse Events Associated with Da Vinci Xi-Assisted Anatomical Lung Resections: A MAUDE Database Analysis of Lobectomy and Segmentectomy},
  journal = {Circulation},
  year = {2026},
  volume = {154},
  number = {Suppl_1},
  doi = {10.1161/circ.154.suppl_1.4529}
}
Kostenlos starten →