Impact of oxygen delivery location during non-invasive home mechanical ventilation: a randomised cross-over simulation study

Dario Somaini, Zoe Bousraou, Lara Benning, Silvia Ulrich iD, Martin Lüchinger, Esther I Schwarz

Respiration · 14.11.2026

DOI: 10.1159/res/aepag028 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Respiration, online erschienen am 14.11.2026. Verfasst von Dario Somaini und 5 weiteren Autor:innen.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

In non-invasive home mechanical ventilation (NIV) using single-limb circuits with vented masks, supplemental oxygen can be added at different locations, potentially resulting in variable delivered inspired oxygen fractions (FiO₂). Recommendations and data comparing oxygen inlet positions are lacking.

Methods

A lung simulator study was conducted using a 1.8 m single-limb NIV circuit with a vented mask. Oxygen was added at three locations: the ventilator oxygen port, a bleed-in connector near the ventilator outlet, and a bleed-in connector between the tube and mask. Oxygen flow rates of 1, 2, and 3 L/min were tested. Three simulated conditions were studied by adjusting lung compliance and airway resistance: healthy, COPD, and obesity. NIV was applied in pressure-controlled ventilation mode. Settings were EPAP/IPAP 4/14 cmH₂O for the healthy model, and both 4/18 and 8/20 cmH₂O for the COPD and obesity models, with a respiratory rate of 14/min. Mean resulting FiO₂ and minute ventilation from three repeated measurements in random order were recorded.

Results

Oxygen delivery via the ventilator oxygen-port and ventilator-near bleed-in produced broadly similar FiO2 values, ad both generally higher FiO₂ than mask-near delivery. Differences between inlet positions became more pronounced at higher oxygen flows reaching a maximum absolute difference of 5.3 percentage points between oxygen port an mask-near delivery. FiO₂ also differed between the two tested ventilator pressure settings in the COPD and obesity simulations, despite only small differences in minute ventilation. Increasing oxygen flow increased FiO₂ across all configurations, although the magnitude of this effect depended on oxygen-injection location and the simulated condition.

Conclusion

In this lung simulator study using a single-limb NIV circuit with a vented mask, oxygen injection location influenced the achieved FiO₂. Under the tested conditions, oxygen delivery via the ventilator oxygen port and a ventilator-near bleed-in connector resulted in broadly similar FiO₂ values and generally achieved higher FiO₂ than mask-near delivery. These findings warrant validation across different ventilators, circuits, interfaces, and pressure settings.

Autor:innen und Institutionen

ORCID-Profile: Silvia Ulrich

Zitieren

APA

Somaini, D., Bousraou, Z., Benning, L., Ulrich, S., Lüchinger, M., & Schwarz, E. I. (2026). Impact of oxygen delivery location during non-invasive home mechanical ventilation: a randomised cross-over simulation study. Respiration. https://doi.org/10.1159/res/aepag028

Vancouver

Somaini D, Bousraou Z, Benning L, Ulrich S, Lüchinger M, Schwarz EI. Impact of oxygen delivery location during non-invasive home mechanical ventilation: a randomised cross-over simulation study. Respiration. 2026. doi:10.1159/res/aepag028

BibTeX
@article{Somaini2026,
  author = {Somaini, Dario and Bousraou, Zoe and Benning, Lara and Ulrich, Silvia and Lüchinger, Martin and Schwarz, Esther I},
  title = {Impact of oxygen delivery location during non-invasive home mechanical ventilation: a randomised cross-over simulation study},
  journal = {Respiration},
  year = {2026},
  doi = {10.1159/res/aepag028}
}
Kostenlos starten →