Physiological and clinical comparison of two high-frequency oscillatory ventilation with volume guarantee strategies in preterm infants with respiratory distress syndrome
Neonatology · 14.11.2026
DOI: 10.1159/neo/adsag044 ↗Kurz gefasst
Fachartikel in der Zeitschrift Neonatology, online erschienen am 14.11.2026. Verfasst von Kamal Ali und 9 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Themen: Neonatologie
Abstract (Originalsprache)
Objective
To compare the physiological performance and clinical outcomes of two high-frequency oscillatory ventilation with volume guarantee (HFOV-VG) strategies in preterm infants born at less than 30 weeks of gestation.
Methods
In this retrospective cohort study, 88 preterm infants born at less than 30 weeks of gestation received either a low-VThf/high-frequency strategy (oscillatory frequency 15–17 Hz; measured VThf/kg 1.5–1.9 mL/kg; n=61) or a high-VThf/standard-frequency strategy (oscillatory frequency 10–12 Hz; measured VThf/kg 2.0–3.0 mL/kg; n=27). Minute-level ventilator and paired arterial blood gas data were analysed. Clinical outcomes included moderate-to-severe bronchopulmonary dysplasia (BPD), mortality, and duration of invasive mechanical ventilation.
Results
A total of 267,786 ventilator observations and 825 paired blood gas observations were analysed. PaCO₂ was comparable between strategies (50 [44–56] vs 49 [43–57] mmHg; p=0.67). The low-VThf/high-frequency strategy was associated with lower FiO₂ requirements, reduced oscillatory pressure variability, and shorter invasive mechanical ventilation (10 [6–21] vs 24 [12–39] days; p=0.0005). After adjustment for gestational age, antenatal corticosteroid exposure, and baseline oxygenation index, the high-VThf/standard-frequency strategy remained independently associated with longer invasive mechanical ventilation (duration ratio 1.65, 95% CI 1.13–2.41; p=0.009), but not moderate-to-severe BPD.
Conclusions
In preterm infants born at less than 30 weeks of gestation, a low-VThf/high-frequency HFOV-VG strategy maintained comparable arterial PaCO₂ control while reducing respiratory support requirements, supporting its use as a physiologically favourable ventilation strategy.
Zitieren
APA
Ali, K., Homedi, A., Algarni, S., Alshareedah, S., Sufyani, M., Almutairi, O., Alsaif, S., Ali, I., Dassios, T., & Greenough, A. (2026). Physiological and clinical comparison of two high-frequency oscillatory ventilation with volume guarantee strategies in preterm infants with respiratory distress syndrome. Neonatology. https://doi.org/10.1159/neo/adsag044
Vancouver
Ali K, Homedi A, Algarni S, Alshareedah S, Sufyani M, Almutairi O, et al. Physiological and clinical comparison of two high-frequency oscillatory ventilation with volume guarantee strategies in preterm infants with respiratory distress syndrome. Neonatology. 2026. doi:10.1159/neo/adsag044
BibTeX
@article{Ali2026,
author = {Ali, Kamal and Homedi, Abdulaziz and Algarni, Saleh and Alshareedah, Saad and Sufyani, Mohammed and Almutairi, Omar and Alsaif, Saif and Ali, Ibrahim and Dassios, Theodore and Greenough, Anne},
title = {Physiological and clinical comparison of two high-frequency oscillatory ventilation with volume guarantee strategies in preterm infants with respiratory distress syndrome},
journal = {Neonatology},
year = {2026},
doi = {10.1159/neo/adsag044}
}
EUVIMED