Monitoring of Remote Ischemic Conditioning with a Tissue Reflectance Sensor: A Randomized Control Trial
Cerebrovascular Diseases · 10.10.2026
DOI: 10.1159/ced/accag037 ↗Kurz gefasst
Fachartikel in der Zeitschrift Cerebrovascular Diseases, online erschienen am 10.10.2026. Verfasst von Gurpriya Bhangoo und 10 weiteren Autor:innen. Beteiligte Einrichtung: Faculty of Nursing, University of Alberta, Edmonton, AB, Canada. Die Arbeit nennt 48 Literaturangaben. Gefördert von: University Hospital Foundation Medical Research Competition.
Abstract (Originalsprache)
Introduction
Remote ischemic conditioning (RIC) is delivered by a transient (5 min) pressure increase in the arm BP cuff by a variable 30–50 mm Hg above systolic BP to a maximum of 200 mm Hg. This inconsistency in pressure application may affect the fidelity of intervention delivery. We tested the hypothesis that monitoring remote ischemic conditioning (RIC) with an adjunct forearm skin tissue reflectance sensor (STRS) device may be feasible in patients with acute ischemic stroke (AIS) and cerebral small vessel disease (cSVD).
Methods
We screened patients with AIS within 7 days of symptom onset for moderate-to-severe cSVD. Eligible patients were randomized 2:1 to receive intervention RIC or sham RIC. The primary outcome measure was a composite feasibility of the forearm STRS (patient-level comfort and technical aspects measured by IOx1 and IOx2). The secondary outcome was changes in forearm STRS-derived tissue reflectance responses during the pressure-on phase (5 min) and the early pressure-off phase (1 min) of RIC. Multivariable linear regression analysis assessed the relationship between IOx1 and intervention allocation. We also used receiver operating characteristic curve analysis to propose candidate thresholds for classifying the magnitude of the tissue oxygenation response to RIC.
Results
Fifty-one (34 intervention [72 RIC sessions], 17 sham [29 RIC sessions]) patients with mean ± SD age of 69 ± 12.8 years, 24(47.1%) were females, were enrolled at a median (IQR) 42(26–78) hours after symptom onset, and median baseline National Institute of Health stroke Scale of 5(3–7). STRS-derived median relative IOx1(β = 0.33, p = 0.002) and median relative IOx2 (β = 0.23, p = 0.04) slopes during the pressure-on phase were different in the intervention arm compared to the sham arm. A percent IOx1 change of 4–13.7% can be considered a threshold to differentiate between intervention and sham treatment. The median Likert scale score was similar: 3.5 (3–4) in the intervention group and 4 (4–5) in the sham group.
Discussion
RIC monitoring with forearm STRS is feasible in patients with AIS. Clinical trials evaluating the efficacy of RIC interventions can use STRS to assess fidelity. Clinical Trial Registration Number ClinicalTrials.gov ID: NCT05408130 (https://clinicaltrials.gov/study/NCT05408130).
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Faculty of Nursing, University of Alberta, Edmonton, AB, Canada.
- Department of Medicine, University of Alberta, Edmonton, AB, Canada.
- Department of Medicine, Queen's University, Kingston, ON, Canada.
- Department of Medicine, University of Calgary, Calgary, AB, Canada.
- Institute for Smart Augmentative and Restorative Technologies and Health Innovations (iSMART), University of Alberta, Edmonton, AB, Canada.
- Faculty of Rehabilitation Medicine, University of Alberta, Edmonton, AB, Canada.
Förderung
- University Hospital Foundation Medical Research Competition
Zitieren
APA
Bhangoo, G., Kashinkunti, C., Sarmiento, R. J., Khan, N., Nair, R. K., Buck, B., Gauthier, M., Mushahwar, V., Shuaib, A., Ferguson-Pell, M., & Kate, M. (2026). Monitoring of Remote Ischemic Conditioning with a Tissue Reflectance Sensor: A Randomized Control Trial. Cerebrovascular Diseases. https://doi.org/10.1159/ced/accag037
Vancouver
Bhangoo G, Kashinkunti C, Sarmiento RJ, Khan N, Nair RK, Buck B, et al. Monitoring of Remote Ischemic Conditioning with a Tissue Reflectance Sensor: A Randomized Control Trial. Cerebrovascular Diseases. 2026. doi:10.1159/ced/accag037
BibTeX
@article{Bhangoo2026,
author = {Bhangoo, Gurpriya and Kashinkunti, Chetan and Sarmiento, Robert Joseph and Khan, Nabeela and Nair, Radhika Kadangot and Buck, Brian and Gauthier, Michel and Mushahwar, Vivian and Shuaib, Ashfaq and Ferguson-Pell, Martin and Kate, Mahesh},
title = {Monitoring of Remote Ischemic Conditioning with a Tissue Reflectance Sensor: A Randomized Control Trial},
journal = {Cerebrovascular Diseases},
year = {2026},
doi = {10.1159/ced/accag037}
}
EUVIMED