Residual Apnoea-Hypopnoea Index Measurement from Positive Airway Pressure Devices versus Home Polygraphy: A Method Comparison Study Using Real-World Data
Respiration · 25.06.2026 · 1-11
DOI: 10.1159/000553192 ↗Kurz gefasst
Fachartikel in der Zeitschrift Respiration, online erschienen am 25.06.2026. Verfasst von Stephanie K Mansell und 8 weiteren Autor:innen. Beteiligte Einrichtung: University College London, London, UK, stephanie.mansell.21@ucl.ac.uk. Die Arbeit nennt 38 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Introduction
Residual apnoea-hypopnoea index (AHI) derived from positive airway pressure (PAP) device downloads is frequently used to assess treatment effectiveness in obstructive sleep apnoea (OSA). However, the accuracy of device reported AHI in real-world settings remains uncertain. This study evaluated the agreement between PAP device-derived AHI and AHI measured using home-based cardiorespiratory polygraphy.
Methods
This secondary analysis of the 3DPiPPIn randomised controlled trial included adults with OSA (AHI ≥15 events/h) naïve to domiciliary PAP therapy. Participants underwent two-night home cardiorespiratory polygraphy at 3 and 6 months. Residual AHI was obtained from Löwenstein PAP devices via data download. Agreement between methods was assessed using Bland-Altman analysis, intraclass correlation coefficients (ICCs), and weighted Kappa. Sensitivity and specificity for detecting inadequate control (AHI ≥7 events/h) were calculated. The association between average leak and AHI discrepancy was examined using linear regression.
Results
Ninety-eight participants were included. PAP-derived AHI values consistently underestimated polygraphy-derived AHI across all time points. Agreement was poor, with ICCs of 0.02 at 3 months and 0.16 at 6 months and low diagnostic agreement across severity categories at both 3 months (K0.10, 95% CI: 0 to 0.26) and 6 months (K0.09, 95% CI: 0 to 0.28). Bland-Altman plots demonstrated systematic underestimation with wide limits of agreement. Sensitivity for detecting inadequate control was very low at both time points (3 months 12% and 6 months 21%), despite high specificity (97% at 3 months and 95% at 6 months). Higher interface leak was significantly associated with greater underestimation of AHI (3 months: slope = 0.45, R2 = 0.18, p < 0.0001; 6 months: slope = 0.38, R2 = 0.14, p < 0.0001).
Conclusion
PAP device downloads substantially underestimated residual AHI and demonstrated poor agreement with home cardiorespiratory polygraphy. These findings highlight important limitations of device-derived AHI in real-world practice, particularly in the presence of interface leak. Where treatment adequacy is uncertain, confirmatory assessment with polygraphy remains necessary.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- University College London, London, UK, stephanie.mansell.21@ucl.ac.uk.
- Royal Free London NHS Foundation Trust, London, UK, stephanie.mansell.21@ucl.ac.uk.
- Royal Free London NHS Foundation Trust, London, UK.
- University College London, London, UK.
- Department of Health Sciences, Brunel University of London, London, UK.
Zitieren
APA
Mansell, S. K., Augustt, S., Gowing, F., Ridout, D., Hilton, S., Kilbride, C., Main, E., Schievano, S., & Mandal, S. (2026). Residual Apnoea-Hypopnoea Index Measurement from Positive Airway Pressure Devices versus Home Polygraphy: A Method Comparison Study Using Real-World Data. Respiration, 1-11. https://doi.org/10.1159/000553192
Vancouver
Mansell SK, Augustt S, Gowing F, Ridout D, Hilton S, Kilbride C, et al. Residual Apnoea-Hypopnoea Index Measurement from Positive Airway Pressure Devices versus Home Polygraphy: A Method Comparison Study Using Real-World Data. Respiration. 2026:1-11. doi:10.1159/000553192
BibTeX
@article{Mansell2026,
author = {Mansell, Stephanie K and Augustt, Sharon and Gowing, Francesca and Ridout, Deborah and Hilton, Stephen and Kilbride, Cherry and Main, Eleanor and Schievano, Silvia and Mandal, Swapna},
title = {Residual Apnoea-Hypopnoea Index Measurement from Positive Airway Pressure Devices versus Home Polygraphy: A Method Comparison Study Using Real-World Data},
journal = {Respiration},
year = {2026},
pages = {1--11},
doi = {10.1159/000553192}
}
EUVIMED