Fachartikel in der Zeitschrift Psychotherapy and Psychosomatics, online erschienen am 08.06.2026. Verfasst von Bella Bingbing Zhang und 13 weiteren Autor:innen. Beteiligte Einrichtung: Department of Rehabilitation Sciences, The Hong Kong Polytechnic University, Hong Kong, China. Die Arbeit nennt 93 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Themen: Depression
Abstract (Originalsprache)
Introduction
Repetitive transcranial magnetic stimulation (rTMS) is a widely utilized, noninvasive brain stimulation technique for treating neurological and psychiatric disorders worldwide. The efficacy of rTMS is influenced by individual head morphology, which shapes the induced electric field (E-field) in the brain. The standard practice for dosing rTMS involves adjusting the intensity to a defined percentage of the motor threshold (MT), a method exemplified by the globally adopted, US FDA-approved protocol of using 120% MT for depression. Since head morphology varies systematically across ethnic groups, current dosing protocols may lead to inconsistent cortical dosing.
Methods
Here, we used large-scale computational modeling (N = 1,085) to systematically evaluate TMS dosage across “White,” “Han Chinese,” and “Black or African Am.” populations.
Results
Our findings reveal striking ethnic disparities in the E-field ratio between the therapeutic target (left dorsolateral prefrontal cortex, DLPFC) and the calibration site (left primary motor cortex, M1). “Han Chinese” individuals exhibited a significantly higher E-field ratio (E-fieldDLPFC/E-fieldM1 = 1.226) compared to “White” individuals (E-fieldDLPFC/E-fieldM1 = 1.096). These differences are primarily driven by variations in scalp-to-cortex distance (SCD). Based on these findings, we derived both simplified population-level adjustment formulas and individualized SCD-based dosing equations and implemented them in a web-based tool for practical use.
Conclusion
Our results suggest that applying a uniform multiplier of MT to DLPFC stimulation may introduce systematic differences in delivered cortical dose across ethnic populations, highlighting that universal dosing conventions derived predominantly from specific demographic cohorts may embed bias, such that established safety and efficacy profiles may not be uniformly applicable across diverse global populations.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Rehabilitation Sciences, The Hong Kong Polytechnic University, Hong Kong, China.
- Shanghai YangZhi Rehabilitation Hospital (Shanghai Sunshine Rehabilitation Center), School of Medicine, Tongji University, Shanghai, China.
- Shenzhen Institutes of Advanced Technology, Guangdong-Hong Kong-Macao Joint Laboratory of Human-Machine Intelligence-Synergy Systems, Chinese Academy of Sciences, Shenzhen, China.
- Department of Psychiatry and Psychotherapy, LMU University Hospital, LMU Munich, Munich, Germany.
- Max Planck Institute of Psychiatry (MPIP), Munich, Germany.
- Department of Psychiatry, Psychotherapy, and Psychosomatics, Medical Faculty, University of Augsburg, Augsburg, Germany.
Zitieren
APA
Zhang, B. B., Lin, T. T., Qin, P. P., Kan, R. L., Xia, A. W., Jin, M., Peng, Y., Keeser, D., Padberg, F., Tik, M., Yuan, T., Becker, B., Zhu, F. F., & Kranz, G. S. (2026). How Standardized Is Transcranial Magnetic Stimulation Treatment for Depression? Large-Cohort Modeling Reveals Systematic Dosimetric Variability. Psychotherapy and Psychosomatics, 1-17. https://doi.org/10.1159/000552942
Vancouver
Zhang BB, Lin TT, Qin PP, Kan RL, Xia AW, Jin M, et al. How Standardized Is Transcranial Magnetic Stimulation Treatment for Depression? Large-Cohort Modeling Reveals Systematic Dosimetric Variability. Psychotherapy and Psychosomatics. 2026:1-17. doi:10.1159/000552942
BibTeX
@article{Zhang2026,
author = {Zhang, Bella Bingbing and Lin, Tim Tianze and Qin, Penny Ping and Kan, Rebecca Laidi and Xia, Adam Weili and Jin, Minxia and Peng, Yinghu and Keeser, Daniel and Padberg, Frank and Tik, Martin and Yuan, Tifei and Becker, Benjamin and Zhu, Frank Fan and Kranz, Georg S},
title = {How Standardized Is Transcranial Magnetic Stimulation Treatment for Depression? Large-Cohort Modeling Reveals Systematic Dosimetric Variability},
journal = {Psychotherapy and Psychosomatics},
year = {2026},
pages = {1--17},
doi = {10.1159/000552942}
}
EUVIMED